Back to search

UCL (University College London)

A Scintillator-Based Range Telescope for Particle Beam Radiotherapy

Abstract

dc:description.abstract

Particle beam therapy (PBT) is a form of radiation therapy that is used for cancer treatment. Recently, interest in scintillator-based detectors for the measurement of depth-dose curves of therapeutic particle beams has been growing. In this work, a novel range telescope based on plastic scintillator and read out by a large-scale CMOS image sensor is presented. The detector is made of a stack of 49 plastic scintillator sheets with a thickness of 2–3mm and a transverse area of 100 times 100mm2. A novel Bragg curve model that incorporates scintillator quenching effects was developed for the beam range reconstruction from depth-light curves with low depthresolution. Measurements to characterise the performance of the detector were carried out at three different PBT centres across Europe. The maximum difference between the measured proton range and the reference range was found to be 0.46 mm. An evaluation of the radiation hardness proved that the range reconstruction algorithm is robust following the deposition of 6,300 Gy peak dose into the detector. Variations in the beam spot size, the transverse beam position and the beam intensity were shown to have a negligible effect on the range reconstruction accuracy. Range measurements of helium, carbon and oxygen ion beams were also performed. A novel technique for online range verification based on a mixed helium/ carbon ion beam and the range telescope was investigated. The helium beam range modulation by a narrow air gap of 1mm thickness in a plastic phantom that affected less than a quarter of the beam particles was detected, demonstrating the outstanding sensitivity of the mixed-beam technique. Using two anthropomorphic pelvis phantoms it was shown that small rotations of the phantom as well as simulated bowel gas movements cause detectable range changes. The future prospects and limitations of the helium-carbon mixing as well as its technical feasibility are discussed.

Degree

thesis:*
Level dc:type.qualificationlevel
Doctoral
Grantor dc:publisher.institution
UCL (University College London)
Year dc:date.issued
2021

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Kelleter, Laurent
Contributors dc:contributor
  • Jolly, S

Rights

Language dc:language
en, eng

Identifiers

dc:identifier.*
Repository record dc:identifier.uri
https://discovery.ucl.ac.uk/id/eprint/10119396/

Chain of custody

source
Harvested from
University College London
Base URL
discovery.ucl.ac.uk/cgi/oai2
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
related terms
citation

Kelleter, Laurent. A Scintillator-Based Range Telescope for Particle Beam Radiotherapy. Doctoral thesis, UCL (University College London), 2021. https://discovery.ucl.ac.uk/id/eprint/10119396/