Electrochemical safety of neural implants: in vitro study of platinum electrodes behaviour near Shannon’s safe stimulation limit

Abstract

dc:description.abstract

In neural electrical stimulation, safe stimulation guidelines are crucial for effective treatment while avoiding neural damage and electrode degradation. Shannon’s parameter k sets conditions on stimulation parameters to prevent neural damage, though the underlying mechanisms remain unclear. An experimental approach was designed to explore the electrochemical mechanisms involved in the charge injection behaviour of platinum electrodes near Shannon’s safe limit. A custom stimulator was developed to deliver biphasic, cathodic-first, charge-balanced, asymmetric, current-controlled pulse trains, while monitoring electrode potential. The pulse trains were calibrated to approach and exceed Shannon’s limit. A characterisation and conditioning sequence, including electrochemical impedance spectroscopy and cyclic voltammetry, was implemented to ensure repeatability. The influence of various experimental parameters was investigated, such as electrode design (recess, electrode size, electrode shape), stimulation protocol (pulse width), and electrolyte composition (pH, ionic strength, density, buffer system, dissolved gas). Initial results in saline confirmed previous findings: positive potential ratcheting from irreversible charge injection linked to molecular oxygen reduction, and correlation of Shannon’s limit with oxide-mediated platinum dissolution. This mechanism held consistent despite parameter variations, though the onset k-value differed. Changes in electrode design affected current density distribution and diffusion profile; variations in pulse width influenced polarisation due to longer diffusion times; electrolyte composition changes impacted ionic interactions, including double-layer capacitance, faradaic mechanisms, and diffusion. Double-layer capacitance and diffusion profile emerged as key factors determining oxidemediated platinum dissolution threshold and rate. Future research should further examine how the in vivo environment impacts these parameters and work towards developing a standardised characterisation method for stimulation electrodes.

Degree

thesis:*
Level dc:type.qualificationlevel
Doctoral
Grantor dc:publisher.institution
UCL (University College London)
Year dc:date.issued
2024

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Niederhoffer, Thomas
Contributors dc:contributor
  • Lancashire, Henry
  • Vanhoestenberghe, Anne

Rights

Language dc:language
en, eng

Identifiers

dc:identifier.*
Repository record dc:identifier.uri
https://discovery.ucl.ac.uk/id/eprint/10199947/

Chain of custody

source
Harvested from
University College London
Base URL
discovery.ucl.ac.uk/cgi/oai2
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
related terms
citation

Niederhoffer, Thomas. Electrochemical safety of neural implants: in vitro study of platinum electrodes behaviour near Shannon’s safe stimulation limit. Doctoral thesis, UCL (University College London), 2024. https://discovery.ucl.ac.uk/id/eprint/10199947/