Back to search

University of North Texas

Microcantilever Based Viscosity Measurement as it Applies to Oscillation Amplitude Response

Abstract

dc:description

The goal of this research is to measure viscosity via the analysis of amplitude response of a piezo driven vibrating cantilevers partially immersed in a viscous medium. As a driving frequency is applied to a piezoceramic material, the external forces acting on the system will affect its maximum amplitude. This thesis applies this principle through experimental and analytical analyses of the proportional relationship between viscosity and the amplitude response of the first natural frequency mode of the sinusoidal vibration. Currently, the few cantilever-based viscometer designs that exist employ resonant frequency response as the parameter by which the viscosity is correlated. The proposed piezoelectric viscometer employs amplitude response in lieu of resonant frequency response. The goal of this aspect of the research was to provide data confirming amplitude response as a viable method for determining viscosity. A miniature piezoelectric plate was mounted to a small stainless-steel cantilever beam. The tip of the cantilever was immersed within various fluid test samples. The cantilever was then swept through a range of frequencies in which the first frequency mode resided. The operating principle being as the viscosity of the fluid increases the amplitude response of cantilever vibration will decrease relatively. What was found was in fact an inversely exponential relationship between dynamic viscosity and the cantilever beam's vibrational amplitude response. The experiment was performed using three types of cantilevers as to experimentally test the sensitivity of each.

Degree

thesis:*
Grantor dc:publisher
University of North Texas
Year dc:date
2018

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Siegel, Sanford H.
Contributors dc:contributor
  • Zhang, Haifeng
  • Choi, Tae-Youl
  • Li, Xiaohua

Subjects

dc:subject × 7

Rights

dc:rights
Statement dc:rights
  • Public
  • Siegel, Sanford H.
  • Copyright
  • Copyright is held by the author, unless otherwise noted. All rights Reserved.
Language dc:language
English

Identifiers

dc:identifier.*
Identifier
local-cont-no: submission_1305
https://digital.library.unt.edu/ark:/67531/metadc1248454/
ark: ark:/67531/metadc1248454

Chain of custody

source
Harvested from
University of North Texas
Base URL
digital.library.unt.edu/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Siegel, Sanford H.. Microcantilever Based Viscosity Measurement as it Applies to Oscillation Amplitude Response. University of North Texas, 2018. https://doi.org/10.12794/metadc1248454