Abstract
dc:descriptionNumerical simulation is used to investigate the accuracy of a direct-contact device for measuring skin-surface temperature. A variation of thermal conductivity of the foam has greater effect on the error rather than a variation of the blood perfusion rate. For a thermal conductivity of zero, an error of 1.5 oC in temperature was identified. For foam pad conductivities of 0.03 and 0.06 W/m-oC, the errors are 0.5 and 0.15 oC. For the transient study, with k=0 W/m-oC, it takes 4,900 seconds for the temperature to reach steady state compared with k=0.03 W/m-oC and k=0.06 W/m-oC where it takes 3,000 seconds. The configuration without the foam and in presence of an air gap between the skin surface and the sensor gives the most uniform temperature profile.
Degree
thesis:*- Grantor dc:publisher
- University of North Texas
- Year dc:date
- 2008
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Dugay, Murielle
- Contributors dc:contributor
-
- Boetcher, Sandra
- Traum, Matthew J.
- Boubekri, Nourredine
Subjects
dc:subject × 7Rights
dc:rights- Statement dc:rights
-
- Public
- Copyright
- Dugay, Murielle
- Copyright is held by the author, unless otherwise noted. All rights reserved.
- Language dc:language
- English
Identifiers
dc:identifier.*- Identifier
-
oclc: 437349288
https://digital.library.unt.edu/ark:/67531/metadc9786/
ark: ark:/67531/metadc9786