Numerical simulation is used to investigate the accuracy of a direct-contact device for measuring skin-surface temperature. A variation of thermal conductivity of the foam has greater effect on the error rather than a variation of the blood perfusion rate. For a thermal conductivity of zero, an error of 1.5 oC in temperature was identified. For foam pad conductivities of 0.03 and 0.06 W/m-oC, the errors are 0.5 and 0.15 oC. For the transient study, with k=0 W/m-oC, it takes 4,900 seconds for the temperature to reach steady state compared with k=0.03 W/m-oC and k=0.06 W/m-oC where it takes 3,000 seconds. The configuration without the foam and in presence of an air gap between the skin surface and the sensor gives the most uniform temperature profile.

Degree

thesis:*
Grantor dc:publisher
University of North Texas
Year dc:date
2008

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Dugay, Murielle
Contributors dc:contributor
  • Boetcher, Sandra
  • Traum, Matthew J.
  • Boubekri, Nourredine

Subjects

dc:subject × 7

Rights

dc:rights
Statement dc:rights
  • Public
  • Copyright
  • Dugay, Murielle
  • Copyright is held by the author, unless otherwise noted. All rights reserved.
Language dc:language
English

Identifiers

dc:identifier.*
Identifier
oclc: 437349288
https://digital.library.unt.edu/ark:/67531/metadc9786/
ark: ark:/67531/metadc9786

Chain of custody

source
Harvested from
University of North Texas
Base URL
digital.library.unt.edu/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Dugay, Murielle. Errors in skin temperature measurements.. University of North Texas, 2008. https://doi.org/10.12794/metadc9786