Back to search

Virginia Commonwealth University

REAL TIME 3-D TRACKING OF THE HIGH DOSE RATE RADIATION SOURCE USING A FLAT PANEL DETECTOR

Abstract

dc:description.abstract

A number of QA procedures have been developed for Breast Brachytherapy treatments, yet none guarantee accurate dose delivery or allow conformation of the actual source position leading to errors sometimes going unnoticed. The objective of this study is to track the exact path the HDR source would follow in real time. The exit radiation of the HDR source was used to image a well defined matrix of markers. The images were acquired using FPD and were processed to obtain projection coordinates while an x-ray calibration image was processed to obtain marker coordinates. Each marker along with its projection represents a line in 3D. A mathematical solution for the ‘near-intersection’ of two 3D lines was implemented and used to determine the ‘true’ 3D source position. A matrix with N markers will produce N*(N-1)/2 points of intersection and their mean will result in a more accurate source position. This study has proved that the accuracy of source position detection using a FPD is sub-millimeter.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Biomedical Engineering

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Bondal, Aditya
Contributors dc:contributor
  • Dorin Todor
  • Ding-Yu Fei

Subjects

dc:subject × 6

Rights

dc:rights
Statement dc:rights
  • © The Author

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
Virginia Commonwealth University
Base URL
scholarscompass.vcu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Bondal, Aditya. REAL TIME 3-D TRACKING OF THE HIGH DOSE RATE RADIATION SOURCE USING A FLAT PANEL DETECTOR. Thesis thesis, https://doi.org/10.25772/DZ74-7T53