Abstract
dc:descriptionComparison of the Approaches to Estimate Limit of Quantification in Liquid Chromatography Limit of quantification (LOQ) is perhaps the most important performance characteristic in analytical method validation. In the present work three methods based on the signal-to-noise ratio, linear regression and Eurachem guidelines were compared to estimate the limits of quantification for two parabens using high performance liquid chromatography. Both, instrumental and method LOQs were estimated and compared. The results indicated that the measured LOQs were not equal and varied by a factor of 10 for the different techniques. The lowest LOQ values (0.12-0.17 mg/L) were obtained by linear regression method, the highest (0.81-1.72 mg/L) – based on Eurachem guidelines. It is therefore important to have clearly described procedure for measuring the LOQ during validation to allow interlaboratory comparison. In our opinion, the best approach was a methodology based on signal-to noise ratio.
Degree
thesis:*- Grantor dc:publisher
- Institutional Repository of Vilnius University
- Year dc:date
- 2016
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Choromanskis, Andrius,
- Contributors dc:contributor
-
- Padarauskas, Audrius
Rights
dc:rights- Statement dc:rights
-
- info:eu-repo/semantics/openAccess
- Language dc:language
- lit
Identifiers
dc:identifier.*- Repository record dc:identifier
- https://repository.vu.lt/VU:ELABAETD20217431&prefLang=en_US