dc:description.abstractThis work proposes a new time-domain integration method to realize Electrochemical Impedance Spectroscopy (EIS). Unlike traditional EIS systems using Fast Fourier Transform (FFT) or Frequency Response Analyzer (FRA) based on a clean quadrature sinusoid stimulus, A single low-frequency, low-amplitude sinusoid stimulus is proposed by using a sinusoid DAC which fully eliminates the analog filter. The response current is integrated by a switched capacitor circuit which is synchronized by the proposed sinusoid DAC. The integration output is sampled and digitized by an 8-bit SAR ADC. The phase and amplitude information of target impedance can be extracted from the time-domain data. The error caused by harmonic generation in the sinusoid DAC can be suppressed by the time-domain integration. The (1×1.1) mm^2 prototype is fabricated in a 130nm CMOS process. It consumes 10µA from a 1.2V supply. Compared with the existing EIS systems, the test results show that the system can obtain a reasonable error with much smaller area and much lower power consumption.
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Gu, Jingren
- Advisor dc:contributor.advisor
-
- Otis, Brian
Subjects
dc:subject × 3Rights
dc:rights- Statement dc:rights
-
- Copyright is held by the individual authors.
- Language dc:language.iso
- en_US
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/1773/26148