Back to search

University of Leeds

Mechanical Design Optimisation of a Tibial Insert for use in Instrumented Total Knee Replacement

Abstract

dc:description.abstract

This PhD research has presented guidelines on how to design, manufacture, and test a smart knee implant. The aim of this PhD research was to investigate how to modify a conventional tibial insert design to incorporate sensors and electronic components whilst ensuring patient safety. Several manufacturing methods were developed to address manufacturing and assembling challenges. Computational and experimental studies were carried out to analyse the structural strength, wear performance, and safety of a modified tibial insert. Additional experimental studies were performed to illustrate the feasibility and practical applications for making measurements with embedded sensors. Previous studies on instrumented tibial inserts did not address manufacturing limitations nor safety concerns.

Degree

thesis:*
Name dc:type.qualificationname
Integrated Ph.D and Master
Level dc:type.qualificationlevel
doctoral
Grantor dc:publisher.institution
University of Leeds
Year dc:date.issued
2023

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Kelmers, Edgars
Advisors dc:contributor.advisor
  • Pandit, Hemant
  • van Duren, Bernard H.
  • Jennings, Louise M.
  • Wilcox, Ruth K.
ORCID person:identifier.orcid

Rights

Language dc:language.dcterms_iso639_2
eng

Identifiers

dc:identifier.*
Dc Identifier Grantnumber
EP/L014823/1

Chain of custody

source
Harvested from
White Rose University Consortium
Base URL
etheses.whiterose.ac.uk/cgi/oai2
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
related terms
citation

Kelmers, Edgars. Mechanical Design Optimisation of a Tibial Insert for use in Instrumented Total Knee Replacement. doctoral thesis, University of Leeds, 2023. https://etheses.whiterose.ac.uk/id/eprint/33871/2/Video%201%2C%20CPRESS%2C%20original%20unmodified%20tibial%20insert%2C%20GMK%20Primary%20size%203%20left.avi