• Refine Query
  • Source
  • Publication year
  • to
  • Language
  • 2
  • 1
  • Tagged with
  • 3
  • 3
  • 3
  • 3
  • 3
  • 3
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • 1
  • 1
  • 1
  • About
  • The Global ETD Search service is a free service for researchers to find electronic theses and dissertations. This service is provided by the Networked Digital Library of Theses and Dissertations.
    Our metadata is collected from universities around the world. If you manage a university/consortium/country archive and want to be added, details can be found on the NDLTD website.
1

Implementierung der akustischen Levitation in ein Totalanalysesystem

Warschat, Carsten 20 September 2018 (has links)
Als Totalanalysesysteme (TAS) werden Geräte bezeichnet, welche komplette chemische Analysen eigenständig ausführen. Die Einführung solcher Systeme ermöglicht einen effizienteren Arbeitsablauf in Analyselaboren, da beispielsweise die Probenmanipulation, Aufreinigung und die physikalisch-/chemische Analyse automatisiert in einem Arbeitsgang durchgeführt werden können. Die speziellen Mikrototalanalysesysteme benötigen geringere Probemengen im $\mu$L- Bereich. Durch Kontamination, Agglomeration oder einem Verschluss etwaiger Kanäle in mikrofluidischen Totalanalysesystemen kann es zu einem kompletten Systemausfall kommen. Eine Alternative bildet die akustische Levitation, um derartige Störfälle durch gänzlichen Verzicht auf Gefäße und Wandkontakte gezielt zu reduzieren. Damit die akustische Levitation erfolgreich in Mikrototalanalysesystemen Anwendung finden kann, bedarf es der technischen Weiterentwicklung von Analysemethoden und Kopplungstechniken. In der vorliegenden Arbeit wird das Hauptaugenmerk auf die Kopplung von Levitationstechnik und Massenspektrometrie gelegt. Darüber hinaus wurden spektroskopische Experimente durchgeführt, welche auf Totalreflektionen innerhalb der Tropfen beruhen. Die besonders gute Reflektion hängt damit zusammen, dass sich die Phasengrenze zwischen Luft und Flüssigkeit im Schwebezustand durch molekulare Wechselwirkungen ständig erneuert und keine produktionsbedingte raue Oberfläche aufweist. Die Kombination aus automatischer Tropfengenerierung, Spektroskopie sowie der entwickelten Methode zur Ionenerzeugung aus dem Probevolumen und der massenspektrometrischen Analyse bilden die Grundlage eines neuartigen Mikrototalanalysesystems für geringe Probemengen. / As a total analysis system (TAS) an instrument is called which carries out complete chemical analysis procedures independently. The introduction of such systems offers a more efficient workflow in analytical laboratories because the sample manipulation, purification and the actual automated analysis can be carried out in one single operation. Specialized and already existing micro total analysis systems require currently a small amount of sample in the $\mu$L range. Owing to contamination, agglomeration and thus cross-secion reduction of incorporated channels in micro fluidics total analysis systems it can lead to a complete system interruption. Hence, the implementation of acoustic levitation in these systems is interessting alternative in order to avoid such kind of problems by abandoning vessels and wall contacts completely. To ensure acoustic levitation in micro total analysis systems can be successfully applied, technical development of analytical methods and coupling techniques is required. In the present work, the coupling of levitation technology and mass spectrometry is the prioritized topic but, in addition, spectroscopic experiments based on total reflections within the levitated droplet are as well realized in order to gain process insights. The particularly good reflection at the freely levitated droplet's circumference is due to the fact that the phase boundary between air and liquid is renewed by molecular interactions constantly and has no production-related rough surface. The combination of automated droplet generation, spectroscopy as well as the developed method for ion generation from the sample volume and mass spectrometry forms the basis of a novel micro total analysis system for small sample quantities.
2

Probing levitated droplets with mass spectrometry

Stindt, Arne 30 May 2016 (has links)
Ultraschalllevitation kombiniert die Vorteile von Mikrofluidik, wie beispielsweise die sehr geringe benötigte Probenmenge, mit einer wandlosen Probenhandhabung. Obwohl die Kopplung zwischen le- vitierten Tröpchen und verschiedenster analytischer Methoden wie optischer Spektroskopie und Röntgenbeugung sehr genau untersucht ist, fehlt es immer noch an einer etablierten Kopplung mit einer massenspektrometrischen Methode für die Analyse auf molekularer Ebene. Die vorliegende Arbeit beschreibt die Prinzipien, auf denen eine kontaktlose massenspektrometrische Analyse von levitierten flüssi- gen Proben beruht. Zuerst wurde der neu entworfene akustische Levitator bezüglich des Einflusses seiner Geometrie auf die Levi- tationseigenschaften experimentell und mittels numerischer Simul- tationen untersucht. Die anschließend durch geführten Experimen- te demonstrieren das Potential von Infrarot-Lasern als kombinierte Desorptions- und Ionisationsquelle für organische Substanzen aus einer Mischung aus Wasser und Glycerin als Cromophor. Um einen tieferen Einblick in die hierbei ablaufenden Ionisationsmechanismen zu erhalten, wurde als Modell ein “Sonic-Spray” Konus räumlich per Massenspektrometrie und Laser-induzierter Fluoreszenz untersucht. Levitator-Geometrie auf die Levitationseigenschaften stimmen sehr gut mit numerischen Simulationen überein. Als komplementäre Ionisationsmethode wurde eine Niedertemperatur-Plasmaquelle ein- gesetzt. Nach einer zeitaufgelösten Untersuchung der grundlegenden Ionisationsmechanismen wurde diese Quelle für die Untersuchung flüchtiger Spezies aus der levitierten Probe in deren direkten Umgebung ohne störende Interferenzen ge- nutzt. / Ultrasonic levitation combines advantages of microfluidics like the required small sample volumes with a wall-less sample handling. While the coupling of analytical methods like optical spectroscopy as well as x-ray scattering are very well elaborated, an established mass spectrometric method to obtain molecular analytical information is still lacking. The herein presented work describes the fundamental processes for a contactless mass spectrometric analysis of levitated droplets. First, the influences of the specially designed levitator geometry on the levitation capabilities is described. During further experiments, the use of infrared lasers has proven useful as a combined desorption and ionization source for organic molecules from a mixture of water and glycerol as chromophore. Subsequently, sonic-spray ionization was used to gain a deeper understanding of the ionization processes occurring within the spray plume. Mass spectrometric mapping as well as laser-induced fluorescence were performed to investigate the ionization during an aerodynamic breakup of the micro droplets in the spray process. As a complementary sampling method, the ionization with a low- temperature plasma source is described. First, a time-resolved mass spectrometric investigation of the ionization process is shown. Sub- sequent to this fundamental study, the application of such a plasma source for the direct analysis of volatile compounds from within the droplets in the surrounding environment without interferences from the droplets bulk phase is described.
3

Optische Spektroskopie in der Ultraschallfalle

Schenk, Jonas 16 October 2014 (has links)
Die Kopplung von akustischer Levitation mit optischer Spektroskopie ermöglicht die Untersuchung von Reaktionen und intermolekularen Wechselwirkungen unter besonderen Bedingungen. Mit akustischer Levitation können Proben im µL-Bereich kontaktfrei gehaltert und Verunreinigungen durch Oberflächenkontaminationen ausgeschlossen werden. Durch die Verwendung verschiedener Methoden der optischen Spektroskopie wie Raman-Streuung, UV/Vis-Absorptions- und Fluoreszenz-Spektroskopie konnten die levitierten Proben untersucht werden. Die durchgeführten Untersuchungen zeigen, dass die Ultraschallfalle aufgrund der Wandlosigkeit sowie des Kontakts der levitierten Probe mit der Atmosphäre und der daraus resultierenden möglichen Aufkonzentrierung eine interessante Möglichkeit zur Probenhandhabung in der Mikrofluidik darstellt. Anhand zweier sonochemischer Reaktionen wurde mit Absorptionsspektroskopie gezeigt, dass das Ultraschallfeld des Levitators nur einen sehr geringen Einfluss auf die levitierte Probe ausübt. Mittels Absorptions- und Fluoreszenzspektroskopie wurde die durch die Verdampfung induzierte Aggregation eines Farbstoffes untersucht. Zudem wurde die relative Quantenausbeute des Farbstoffes im levitierten Tropfen bestimmt. Der Kontakt des levitierten Tropfens mit der umgebenden Atmosphäre kann auch zur Aufnahme von Substanzen aus der Atmosphäre führen. Dieser Effekt wurde anhand von levitierten ionischen Flüssigkeiten volumetrisch und schwingungsspektroskopisch bei verschiedenen Luftfeuchten untersucht und die Wechselwirkungen mit dem absorbierten Wasser detailliert charakterisiert. Die Kopplung und simultane Messung von Raman-Streuung und UV/Vis-Spektroskopie ermöglichte die Untersuchung der Bildung und Aggregation von Silber-Nanopartikeln und deren Auswirkungen auf die oberflächenverstärkte Raman-Streuung. Zudem wurde die Stabilisierung von Silber-Nanopartikeln in ionischen Flüssigkeiten und die Wechselwirkungen der Partikel mit den ionischen Flüssigkeiten untersucht. / Acoustic levitation in combination with optical spectroscopy allows for investigations of reactions and intermolecular interactions under specific conditions. Samples with microliter volumes can be handled without contact to solid surfaces resulting in the absence of impurities from surface contamination. Using different optical spectroscopy methods such as Raman, UV/vis, and fluorescence spectroscopy, different levitated samples were characterized in detail within this work. The investigations show that the acoustic levitator is an interesting tool for sample handling due to the wall-less fixture and because of the interaction of the levitated sample with the surrounding atmosphere. This interaction leads to an increase of a solved analyte upon the evaporation of the solvent. The ultrasonic field inside the trap was characterized by investigations of sonochemical reactions, which show a negligible influence of the ultrasonic field on the levitated sample. Absorption as well as fluorescence spectroscopy was used to study the aggregation of a dye due to an increasing concentration upon evaporation of the solvent. Furthermore, the relative quantum efficiency of the dye was determined from levitated droplet experiments. The interaction of the levitated droplet with the surrounding atmosphere can also lead to absorption of airborne substances. This effect was investigated for different levitated ionic liquids. Volumetric and vibrational studies were performed on levitated ionic liquids under different atmospheric humidity conditions to characterize the interaction of ionic liquids with water. Hyphenation of Raman scattering and UV/vis absorption spectroscopy enabled the investigation of the formation and aggregation of silver nanoparticles and the correlation of this information with the recorded surface-enhanced Raman spectra. In addition, the stability of silver nanoparticles in ionic liquids and the interactions of the particles with the ionic liquids were examined.

Page generated in 0.1271 seconds