Avaliação de dose em procedimentos especiais de fluoroscopia : histerossalpingografia e dacriocistografiaLOPES, Cintya Carolina Barbosa January 2006 (has links)
Made available in DSpace on 2014-06-12T23:16:56Z (GMT). No. of bitstreams: 2
arquivo9089_1.pdf: 1085916 bytes, checksum: 99db33e1bf6f9c3487f83df708df47c4 (MD5)
license.txt: 1748 bytes, checksum: 8a4605be74aa9ea9d79846c1fba20a33 (MD5)
Previous issue date: 2006 / Dentre os procedimentos especiais de fluoroscopia destacam-se a histerossalpingografia (HSG) e a dacriocistografia (DCG). A HSG é uma técnica radiodiagnostica capaz de detectar patologias uterinas e tubárias, e é fundamental para a investigação de infertilidade. A DCG trata-se do exame radiográfico do sistema lacrimal, sendo importante para mostrar o nível de obstrução, a presença de dilatação do saco lacrimal, bem como alterações em estruturas vizinhas.
Neste trabalho, foi efetuado o estudo da dose de entrada na pele do paciente devido a estes dois procedimentos especiais de fluoroscopia, bem como a avaliação da dose na equipe médica que executa os exames. Para tanto, foram avaliados os procedimentos de 22 pacientes de HSG e 8 pacientes de DCG. Dosímetros de TLD-100 foram utilizados e fixados na pele dos pacientes em pontos anatômicos envolvidos em cada exame.
No caso da HSG os resultados mostraram que a dose na entrada da pele variou de 0,5 mGy a 73,4 mGy, com um valor médio de 22,1 mGy. A dose no útero foi estimada em 5,5 mGy, e 6,6 mGy foi a dose média estimada para os ovários.
As doses de entrada na pele dos pacientes submetidos a exames de DCG variaram de 2,1 mGy a 10,6 mGy, e a dose média entre os olhos foi de 6,1 mGy.
Os resultados das doses ocupacionais mostraram que, na HSG, a dose média na mão direita do médico é de 4,3 mGy por exame. Este valor é devido ao fato que o médico introduz o meio de contraste manualmente durante as exposições com contraste na histerossalpingografia.
Em relação a DCG, os valores de doses ocupacionais foram da ordem dos valores ambientais, evidenciando que, dentro dos limites permitidos, não há risco para os médicos neste procedimento
|
2 |
Determinação de grandezas dosimétricas de interesse em mamografia usando detectores termoluminescentes / Determination of dosimetric quantities of interest in Mammography using thermoluminescent detectors.Mendoza, Raul Ernesto Camargo 10 February 2010 (has links)
Os órgãos de saúde internacionais e nacionais, como o Ministério da Saúde na portaria 453/98 da Vigilância Sanitária, exigem que a Dose de Entrada na Pele seja avaliada para cada equipamento mamográfico através da leitura de um sistema câmara de ionização-eletrómetro corrigida pelo fator de retroespalhamento. Ao não existir menção explícita na portaría de valores utilizáveis para o fator de retroespalhamento, este trabalho visa à determinação experimental do fator de retroespalhamento, através da utilização dos dosímetros termoluminescentes TLD-100. No estudo são verificadas as dependências geométricas e espectrais do fator de retroespalhamento, assim como do valor da Dose de Entrada na Pele, e da Dose em Profundidade, correspondentes com as técnicas radiográficas empregadas nos exames mamográficos convencionais de rotina. Foram avaliados feixes na faixa de 0,35 mmAl até 0,43 mmAl, tensões do tubo de 25kV, 28kV, 30kV, e 32kV, assim como os três tamanhos de campo disponíveis no Mamógrafo Senographe DMR utilizado, e distancias focofilme iguais a 56cm, 61cm e 66cm. Os resultados obtidos foram comparados com publicações existentes, as quais apresentam resultados obtidos através de Simulação Monte Carlo, câmaras de ionização, e dosímetros TLD-100. Os resultados obtidos neste trabalho permitem estabelecer e discutir as dependências das grandezas dosimétricas estudadas com a Camada Semi-Redutora, tensão do tubo, combinação ânodo-filtro, tamanho de campo, distância foco-filme e espessura da mama. / National and international health organizations such as the Brazilian Ministry of Health, through its Secretary of Health Surveillance establishes in the publication Nº 453/98 that in all mammographic equipments must be evaluated the entrance-skin dose through the readings of an ionization chamber-electrometer system corrected by the backscatter factor, among others factors. Nevertheless, there is no explicit mention for useful values of backscatter factor in this document; the main aim of this work is the experimental determination of backscatter factor through the use of TLD-100 dosimeters. In this study, the geometric and spectral dependencies of the backscatter factor, entrance-skin dose and the in-depth dose were evaluated, corresponding to the most radiographic techniques employed in conventional mammographic procedures, i.e., beam qualities in the range of 0.35 mmAl to 0.43 mmAl, tube voltages from 25kV to 32kV, focus-film distances from 56cm to 66cm, and three field sizes were evaluated. Our results were compared with those previously published obtained through Monte Carlo simulation, ionization chambers and TLD dosimeters. The results obtained in this work allow studying the dependency of the mentioned dosimetric quantities with the half-value layer, tube voltage, anode-filter combination, field size, focusfilm distance and breasting thickness of the breast.
|
3 |
Determinação de grandezas dosimétricas de interesse em mamografia usando detectores termoluminescentes / Determination of dosimetric quantities of interest in Mammography using thermoluminescent detectors.Raul Ernesto Camargo Mendoza 10 February 2010 (has links)
Os órgãos de saúde internacionais e nacionais, como o Ministério da Saúde na portaria 453/98 da Vigilância Sanitária, exigem que a Dose de Entrada na Pele seja avaliada para cada equipamento mamográfico através da leitura de um sistema câmara de ionização-eletrómetro corrigida pelo fator de retroespalhamento. Ao não existir menção explícita na portaría de valores utilizáveis para o fator de retroespalhamento, este trabalho visa à determinação experimental do fator de retroespalhamento, através da utilização dos dosímetros termoluminescentes TLD-100. No estudo são verificadas as dependências geométricas e espectrais do fator de retroespalhamento, assim como do valor da Dose de Entrada na Pele, e da Dose em Profundidade, correspondentes com as técnicas radiográficas empregadas nos exames mamográficos convencionais de rotina. Foram avaliados feixes na faixa de 0,35 mmAl até 0,43 mmAl, tensões do tubo de 25kV, 28kV, 30kV, e 32kV, assim como os três tamanhos de campo disponíveis no Mamógrafo Senographe DMR utilizado, e distancias focofilme iguais a 56cm, 61cm e 66cm. Os resultados obtidos foram comparados com publicações existentes, as quais apresentam resultados obtidos através de Simulação Monte Carlo, câmaras de ionização, e dosímetros TLD-100. Os resultados obtidos neste trabalho permitem estabelecer e discutir as dependências das grandezas dosimétricas estudadas com a Camada Semi-Redutora, tensão do tubo, combinação ânodo-filtro, tamanho de campo, distância foco-filme e espessura da mama. / National and international health organizations such as the Brazilian Ministry of Health, through its Secretary of Health Surveillance establishes in the publication Nº 453/98 that in all mammographic equipments must be evaluated the entrance-skin dose through the readings of an ionization chamber-electrometer system corrected by the backscatter factor, among others factors. Nevertheless, there is no explicit mention for useful values of backscatter factor in this document; the main aim of this work is the experimental determination of backscatter factor through the use of TLD-100 dosimeters. In this study, the geometric and spectral dependencies of the backscatter factor, entrance-skin dose and the in-depth dose were evaluated, corresponding to the most radiographic techniques employed in conventional mammographic procedures, i.e., beam qualities in the range of 0.35 mmAl to 0.43 mmAl, tube voltages from 25kV to 32kV, focus-film distances from 56cm to 66cm, and three field sizes were evaluated. Our results were compared with those previously published obtained through Monte Carlo simulation, ionization chambers and TLD dosimeters. The results obtained in this work allow studying the dependency of the mentioned dosimetric quantities with the half-value layer, tube voltage, anode-filter combination, field size, focusfilm distance and breasting thickness of the breast.
|