Optofluidic Spectroscopy Platform for Detection of HemolysisArchibong, Edikan 20 November 2015 (has links)
In the United States alone, hundreds of millions of blood tests are performed annually, and a significant number of those tests are compromised due to hemolysis: e.g., 31% compromised in emergency rooms (inpatient) and 10% at blood banks, clinics, and other outpatient venues. Currently there is no way to reliably detect hemolysis without plasma separation. As a result, significant delays ensue, potentially negatively affecting patient diagnosis and treatment. In addition to in vitro hemolysis, which compromises the quality of blood tests, hemolysis can also occur in vivo. The in vivo occurrence of hemolysis is an indication of life-threatening complications. Being able to detect early signs of in vivo hemolysis would significantly improve outcomes for many patients, including pregnant women affected by HELLP (Hemolysis, Elevated Liver Enzymes, Low Platelet counts) syndrome. Therefore, there is a critical need to be able to detect hemolysis near the patient, immediately following the collecting of blood sample.
The goal of this research is to provide an alternative to the traditional testing of blood samples, which requires large volumes of blood, centrifugation, and bulky instrumentation. The proposed alternative hemolysis detection system is a simple miniature setup that produces test results in minutes. This miniature, near-patient sensor would improve patients’ diagnosis, treatments, general satisfaction, and overall experience. The potential reduction of healthcare costs associated with hemolysis would be another significant benefit.
The technology demonstrated in this dissertation is based on a novel combination of microfluidics, spectroscopy, and optical-fiber sensing. The microfluidics provide the capability to handle small volumes of liquid and to filter particles from solution. Novel membrane fabrication and modular integration provides the means to characterize and culture the captured particles. Spectroscopy and optical fibers provide the means to characterize the filtrate. These capabilities can be used for not only the detection of hemolysis but also other biomedical applications. .
The first step in detecting hemolysis is to separate blood cells and other unwanted particulates from the plasma needed for optical analysis of concentration of hemoglobin. To that end, we focused initially on the problem of particle separation—specifically, within a microfabricated chamber with a custom-designed transparent membrane. To create a miniature microfluidic system capable of processing microliter blood samples, microelectromechanical systems (MEMS) fabrication techniques were required. The fabrication process included steps such as low-stress vapor deposition, photolithography, plasma, and wet etching. The resulting microdevice proved capable of filtering a variety of biological test fluids, including human lung fibroblast cancer cells from medium. The transparent membrane also allows for spectroscopic studies in broader applications, such as spectroscopic analysis or culturing of the cells retained on the filter. These capabilities were demonstrated using microbeads and cancer cells in solution.
Optical techniques are used to analyze the separated blood plasma for concentration of hemoglobin. To integrate spectroscopic capabilities with the above microfluidics system, an optical fiber–based miniature probe was attached to the microfabricated chamber. As proof of concept, this system was tested in an application that required the measurement of physiologically relevant concentrations of cobalamin (vitamin B12). This application was used to address human error in drug administration showing measurements of cobalamin concentration as an example drug that can be monitored. The clinical means range of concentrations is from 1 µg/ml to 1000 µg/ml. The achieved results showed measurements of concentrations between 1 µg /mL to 5 mg/mL to monitor the physiological range and potential overdose in microliter of volume.
This device has potential for numerous applications, ranging from single cell spectroscopy to measurements of glucose concentrations.
This integrated system was then applied to the detection of hemolysis. The complete system conducts optofluidic spectroscopy with the optical fiber probe connected to the microfabricated chamber, which locally filters out blood cells, and reliably determine amount of free hemoglobin with the need for centrifuging. The utility of the device was demonstrated by its accurate measurement of hemoglobin concentrations in blood plasma.
Finally, to apply the concept of the detection system to clinical condition with a reliable, and low-cost system, especially useful for developing countries, a smartphone-based technology, is proposed. This technology delivers ultra-fast results for the detection of early signs of HELLP syndrome and preeclampsia with the goal to decrease mortality and morbidity. The smartphone-based diagnostics is low cost, high speed of operation together with high accuracy. Detection of 1 mg/dL of free hemoglobin was achieved which is comparable to gold standard assay which are time consuming, difficult to operate and expensive.
This technology, in summary, integrates microfluidics with microfiltration and spectroscopic technology to conveniently separate and characterize blood plasma. The device can also provide important information about other complex biological samples. These measurements require only very small sample volumes.
|
2 |
L'anémie iatrogène chez les enfants aux soins intensifs pédiatriquesFrançois, Tine 12 1900 (has links)
Plus de la moitié des enfants survivant à une maladie critique sont anémiques à leur sortie
des soins intensifs pédiatriques (SIP). Cette proportion est inquiétante puisqu’il existe une
association entre l’anémie et un développement neurocognitif anormal chez le jeune enfant. Les
causes d’anémie aux soins intensifs sont multiples. Le volume de sang prélevé à visée
diagnostique est un facteur de risque probable mais aussi un facteur de risque qui est
potentiellement modifiable.
Ce mémoire explore la contribution des prélèvements sanguins à la prévalence d’anémie
chez les enfants atteints d’une maladie critique. L’objectif global est d’une part d’évaluer les
pratiques cliniques actuelles de prélèvements sanguins et d’autre part d’évaluer les stratégies
étudiées dans la littérature pouvant nous permettre de réduire de manière sécuritaire les
quantités de sang retirées à ces enfants. On présente deux articles : une étude de cohorte
prospective observationnelle des pratiques de prélèvements sanguins dans notre service de soins
intensifs pédiatriques ainsi qu’un scoping review de la littérature des interventions visant à
réduire les pertes sanguines à visée diagnostique.
La 1ère étude souligne le problème d’anémie iatrogène aux soins intensifs pédiatriques.
Cette étude démontre que le volume de sang prélevé pour des raisons diagnostiques est
significatif et égal à environ 5% du volume de sang circulant total. Nous démontrons également
qu’il y a une association entre les prélèvements sanguins et un risque élevé d’anémie à la sortie
des soins intensifs. La 2ème étude résume la littérature actuelle des stratégies efficaces destinées
à réduire le sang prélevé en volume et fréquence, réduire l’anémie et également les besoins en
transfusion. Les systèmes de prélèvement à circuit clos, les tubes de petit volume, et les tests
sanguins réalisés au chevet pour répondre à une question clinique, sont tous des interventions
prometteuses. La modification des habitudes de prescription de l’équipe médicale semble
également essentielle.
Ce mémoire nous permettra de mieux comprendre le problème de l’anémie iatrogène
chez les enfants admis aux soins intensifs. Il est important de continuer à améliorer notre pratique
actuelle et de minimiser le volume de sang prélevé et/ou gaspillé. / More than 50% of pediatric intensive care unit (PICU) survivors are anemic at PICU
discharge. This proportion is worrisome because of the known negative association between
anemia and neurocognitive development of young children. The etiology of critical care anemia
is multifactorial. Diagnostic blood sampling is an iatrogenic contributor to this problem, but it is
also a potentially modifiable factor.
The overall objective of our project is to explore current clinical practice for diagnostic
blood sampling in PICU and to look for solutions to improve patient care. In this thesis, we present
two articles: a prospective observational cohort study, conducted in our PICU, and a scoping
review of the existing literature on interventions to minimize diagnostic blood loss.
The 1st study highlights the existing problem of iatrogenic anemia in PICU. Our study
demonstrates that the volume of blood sampled for diagnostic purposes is significant. Nearly 5%
of total circulating blood volume is removed from a critical ill child during a PICU stay. This study
detected an association between blood sampling volume and a higher risk of anemia at PICU
discharge. The 2nd study summarizes current evidence on efficacious strategies to minimize blood
sampling volume and frequency, anemia, and transfusion. Closed blood sampling devices, small
volume blood collection tubes, point-of-care testing, and medical education, are all promising
interventions.
This thesis, which includes both articles, helps us to better understand the problem of
iatrogenic anemia in critically ill children admitted to the intensive care unit. It is important to
continue to improve daily practice and to minimize blood volume sampled and/or wasted.
|