TeseJosemirAlexandrino.pdf: 4030416 bytes, checksum: 971f2eadd7c4d23fb1945b3a4dfbd1d5 (MD5) / MS, CNPq, FAPESB, SESAB, SECTI e IFBA / O uso massivo de tecnologia na área da saúde tem elevado a efetividade dos procedimentos
médicos, com consequente melhoria da qualidade de vida dos pacientes. Entretanto, tais
tecnologias trazem riscos intrínsecos, que podem resultar em danos, quando mal utilizadas ou
mantidas. Neste contexto, encontram-se os Equipamentos Médico-hospitalares (EMH) que
podem apresentar problemas que resultem em diagnósticos errôneos, terapias inapropriadas
ou que comprometam a segurança dos usuários. A avaliação periódica, e em particular, do
desempenho metrológico, constitui-se em importante meio para alcançar o uso seguro do
EMH. As soluções encontradas para tal avaliação são rigidamente condicionadas às
especificidades de cada tipo de EMH e dificilmente podem ser aproveitadas para outros.
Embora não tenha sido encontrada na literatura uma solução generalista que possa ser
aplicada a determinados tipos de EMH, esta indica viabilidade de aplicação, em função dos
constantes avanços tecnológicos do setor. Neste sentido, este trabalho apresenta uma
metodologia para a avaliação do desempenho metrológico de EMHs e para tal, foi
desenvolvido um sistema que reúne instrumentos de medição integrados a um software
gerenciador de ensaios. Este software realiza a aquisição e monitoramento de dados de
medição relativos às variáveis metrológicas do equipamento sob teste, avalia a adequação
destes dados a requisitos metrológicos preestabelecidos e emite relatório de conformidade. O
sistema permite a configuração de diversos parâmetros, possibilitando a execução de
avaliações de desempenho metrológico em diferentes tipos de EMH. Para fins de validação, o
sistema foi configurado para realizar avaliações em bisturis elétricos e incubadoras neonatais.
Os resultados mostram a eficácia da metodologia desenvolvida e a eficiência da plataforma
computacional para avaliação metrológica de EHM. / The massive use of technology in health care has increased the effectiveness of medical
procedures, with consequent improvement of quality of life of patients. However, such
technologies carry inherent risks that can result in damage when misused or maintained. In
this context, are the medical and hospital equipment (MHE) that can present problems that
result in misdiagnosis, inappropriate therapies or which compromise the safety of users. The
periodic assessment, especially of the metrological performance constitutes an important
means to achieve the MHE safe use. The solutions for such evaluation are tightly constrained
to the specific characteristics of each type of MHE and can hardly be reutilized for others.
Although not found in the literature a general solution that can be applied to certain types of
MHE, this indicates the feasibility of application, in function of the constant technological
advances in the medical equipment industry. Thus, this work presents a methodology for
evaluating the metrological performance of MHE and for such, a system was developed that
brings together measuring instruments integrated into software for testing manager. This
software performs the acquisition and monitoring of measurement data related to the
metrological variables of equipment under test, assesses the adequacy of these data to preestablished
metrological requirements and issues a compliance report. The system allows
configuration of various parameters, enabling the execution of metrological performance
evaluations in different types of MHE. For verification purposes, the system was configured
to perform evaluations of high frequency electrosurgical scalpels and neonatal incubators. The
results show the effectiveness of the developed methodology and the efficiency of the
computational platform for metrological evaluation of MHE.
Identifer | oai:union.ndltd.org:IBICT/oai:192.168.11:11:ri/22633 |
Date | 01 March 2012 |
Creators | Alexandrino, Josemir da Cruz |
Contributors | Lepikson, Herman Augusto, Kalid, Ricardo de Araújo, Hermini, Alexandre Henrique, Kalid, Ricardo de Araújo, Hermini, Alexandre Henrique, Lepikson, Herman Augusto, Ferreira, Cássio Costa, Navarro, Marcus Vinícius Teixeira, Da Silva, José Felício, Magalhães, Robson da Silva |
Publisher | Universidade Federal Da Bahia, Escola Politécnica, Departamento de Engenharia Química, Engenharia industrial, PEI, brasil |
Source Sets | IBICT Brazilian ETDs |
Language | Portuguese |
Detected Language | Portuguese |
Type | info:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/doctoralThesis |
Source | reponame:Repositório Institucional da UFBA, instname:Universidade Federal da Bahia, instacron:UFBA |
Rights | info:eu-repo/semantics/openAccess |
Page generated in 0.0031 seconds