National and Kapodistrian University of Athens
In vivo δοσιμετρία ακτινοθεραπευτικής δέσμης φωτονίων με χρήση a-Si EPID.
Abstract
dc:descriptionPurpose: The aim of this study is the validation of a new easily applicable EPID dosimetry method based on the back projection approach and the comparison of this method with the EBT2 film dosimetry method.Methods: The dosimetric characteristics of the EPID were determined. The methodology to calculate the dose on the central axis of the photon beam was described. The EPID dosimetry method was validated with phantom measurements using an ionization chamber. In vivo measurements of ten prostate patients treated with 3D conformal plans were also performed both with the EPID and the EBT2 films. Results: The phantom measurements revealed a mean dispersion of 1.7% between EPID and ionization chamber doses and 2.2% between EPID and EBT2 film doses. While, the in vivo measurements in prostate patients revealed a mean dispersion of 0.8% between EPID doses and treatment planning calculated doses and 1% between EPID and EBT2 film doses.Conclusions: The presented EPID dosimetry method is accurate, easily applicable to all types of flat panels without the use of any sophisticated software and is not time consuming.
Degree
thesis:*- Grantor dc:publisher
- National and Kapodistrian University of Athens
- Year dc:date
- 2015
Author and committee
dc:creator, dc:contributor.*- Authors dc:creator
-
- Kalliopi, Mandourari
- Μανδουράρη, Καλλιόπη
Subjects
dc:subject × 17Rights
- Language dc:language
- gre
Identifiers
dc:identifier.*- Identifier
- 10.12681/eadd/35911