Back to search

Case Western Reserve University School of Graduate Studies

A Microfabricated Deep Brain Stimulation Electrode

Abstract

dc:description

This work presents a novel 4-sided, 16-channel deep brain stimulation electrode with a flexible extension lead for connectivity with pulse generation electronics. The 3-dimensional electrode enables steering the current field circumferentially. The electrode is fabricated in pieces by microfabrication techniques; the pieces are then assembled mechanically, after which the lead is connected. The electrode is modeled by finite element analysis and tested in vitro to validate the design concept, i.e., targeted stimulation. With monopolar and bipolar configurations, the electric potential in front of the activated site is at least five times larger than that on the other not-activated sides within a 3 mm radius. The resistance of each channel is also measured. Finally, a pull test is performed to evaluate the mechanical strength of the epoxy used to attach the lead to the electrode. The samples withstood a force of 8.66 g at a 60° angle, thus exhibiting high robustness.

Degree

thesis:*
Name thesis:degree_name
Master of Sciences (Engineering)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
EECS - Electrical Engineering
Grantor dc:publisher
Case Western Reserve University School of Graduate Studies
Year dc:date
2009

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Lin, Chia-Hua
Contributors dc:contributor
  • Mehregany, Mehran

Subjects

dc:subject × 8

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Lin, Chia-Hua. A Microfabricated Deep Brain Stimulation Electrode. masters thesis, Case Western Reserve University School of Graduate Studies, 2009. http://rave.ohiolink.edu/etdc/view?acc_num=case1244061398