Back to search

University of Toledo Health Science Campus

IMRT Plan Delivery Verification Utilizing a Spiral Phantom with Radiochromic Film Dosimetry

Abstract

dc:description

The purpose of this study was to develop and report on the implementation and IMRT quality assurance plan delivery verification using a Spiral phantom. The phantom utilizes a cylindrical solid water system with a machined spiral trajectory for insertion of film. Several analyses were performed on various IMRT treatment plans comparing predicted planar fluence dose and measured dose using radiochromic film. A solid water cylindrical IMRT phantom manufactured by GAMMEX which has been machined creating a spiral cavity for placement of radiochromic film (GAFCHROMIC® EBT2) was employed for IMRT plan QA. This spiral phantom is implemented to measure data in a three dimensional (3D) subspace which was not previously demonstrated. The patient treatment plan with predicted planar and volumetric isodose distributions were obtained in Pinnacle treatment planning software (TPS). The patient treatment planning data using a CT data set was then projected onto the spiral phantom in the TPS where planar dose files were generated using a scripting file. The measured data was obtained upon successful delivery of the intended treatment plan with a linear accelerator on the spiral phantom with radiochromic film in place. Comparison of the predicted and measured data provides a quantitative and qualitative assessment and validation of the intended treatment when delivered as planned. The films were subsequently scanned and the measured dose data from the delivered plan were compared with the planar fluence dose maps generated in Pinnacle using RIT113 software. Schematic isodose overlays, vertical and horizontal dose profiles as well as IMRT distance to agreement (DTA) analysis of the predicted and measured dose distributions were shown to be in great accord with one another. One can easily see the value of this phantom as a quantitative and qualitative IMRT plan analysis tool.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Sciences (MSBS)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
College of Medicine
Grantor dc:publisher
University of Toledo Health Science Campus
Year dc:date
2010

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Pichler, Joseph Alan
Chair dc:contributor.committeechair
  • Parsai, E. Ishamael

Subjects

dc:subject × 7

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Pichler, Joseph Alan. IMRT Plan Delivery Verification Utilizing a Spiral Phantom with Radiochromic Film Dosimetry. masters thesis, University of Toledo Health Science Campus, 2010. http://rave.ohiolink.edu/etdc/view?acc_num=mco1288963613