Contaminação Microbiana em Placas de Armazenamento de Fósforo IntrabucaisSouza, Tricia Murielly Pereira de 31 January 2014 (has links)
Submitted by Etelvina Domingos (etelvina.domingos@ufpe.br) on 2015-04-08T19:17:15Z
No. of bitstreams: 2
license_rdf: 1232 bytes, checksum: 66e71c371cc565284e70f40736c94386 (MD5)
DISSERTAÇÃO TRICIA MURIELLY PEREIRA ANDRADE DE SOUZA.pdf: 2322643 bytes, checksum: 4725a9d74b61857b6e6c6446d355c50c (MD5) / Made available in DSpace on 2015-04-08T19:17:15Z (GMT). No. of bitstreams: 2
license_rdf: 1232 bytes, checksum: 66e71c371cc565284e70f40736c94386 (MD5)
DISSERTAÇÃO TRICIA MURIELLY PEREIRA ANDRADE DE SOUZA.pdf: 2322643 bytes, checksum: 4725a9d74b61857b6e6c6446d355c50c (MD5)
Previous issue date: 2014 / O objetivo neste estudo foi analisar contaminação microbiana em placas de
armazenamento de fósforo intrabucais utilizadas em duas clínicas privadas de
Radiologia Odontológica. A amostra foi composta por 50 placas de fósforo: 44 do
grupo teste, sendo 14 da clínica A (sistema Digora Optime®), e 30 da clínica B
(sistema ScanX®); e 6 do grupo controle, sendo 3 placas nunca utilizadas de cada
sistema. Os protocolos de controle de infecção adotados pelas clínicas no uso das
placas de fósforo não incluíam o uso de luvas nas etapas de processamento e
embalagem das placas, mas apresentavam o uso de uma barreira plástica adicional
para a realização de técnicas intrabucais. Para a coleta do material microbiano,
Swabs umedecidos foram esfregados na superfície das placas, e transferidos para
tubos contendo 2ml de solução salina (solução-mãe). Esses tubos foram agitados e
seu conteúdo diluído seriadamente. As diluições foram inoculadas em placas de
Petri contendo Ágar Mueller Hinton (MHA), as quais foram incubadas a 37ºC/48 h
para contagem de Unidades Formadoras de Colônias (UFC). Adicionou-se 1,5 ml da
solução-mãe a 1,5 ml de caldo BHI (Brain Heart Infusion), e incubou-se a 37°C/48 h.
Então, as soluções foram semeadas em MHA, Ágar Mitis Salivarius (MSA), Ágar
Sabouraud Dextrose (SDA), e Ágar Sal Manitol (SMA). Incubou-se a 37°C/48 h, e
observou-se o crescimento microbiano. Na clínica A, 50% das placas de fósforo
apresentaram contaminação, e na clínica B 73,3%. Houve crescimento em SMA
para 100% das placas contaminadas da clínica A, e 86,3% da clínica B, indicando
presença de Staphylococcus sp. Em MSA, houve crescimento para uma das placas
de fósforo da clínica A, sugestivo de Streptococcus sp. Não houve crescimento em
SDA, demonstrando ausência de fungos. Verificou-se crescimento de UFC para
28,5% das placas contaminadas da clínica A (26.4 UFC/placa), e 27,2% da clínica B
(de 26.4 a 80.0 UFC/placa). Não foi observada contaminação nas placas dos
sistemas Scan X e Digora Optime, usadas como controle. Conclui-se que há risco de
contaminação cruzada por placas de armazenamento de fósforo, pois a maioria das
placas analisadas apresentou-se contaminada, principalmente por bactérias do
gênero Staphylococcus. Quantitativamente, essa contaminação ocorreu em baixos
níveis, sendo oriunda da manipulação das placas, já que o uso de uma segunda
barreira plástica diminui o risco de contaminação das placas por microrganismos da
cavidade bucal. Read more
|
2 |
Avaliação do design higiênico de equipamentos que contribuem para a contaminação de carcaças na 1ª etapa de abate de suínos / Evaluation of hygienic design of equipments which contribute to pork cascasses contamination in a slaughterhouseMariot, Roberta Fogliatto January 2010 (has links)
O objetivo deste trabalho foi avaliar a contaminação microbiológica e o isolamento de Salmonella spp. em diferentes estágios da 1° etapa de um abatedouro frigorífico, além de aspectos importantes do design higiênico de equipamentos que demonstraram contribuir para a contaminação das carcaças. Para a avaliação microbiológica das carcaças suínas foram estimadas as contagens de microrganismos heterotróficos totais (CT) e a incidência de Salmonella spp. em quatro locais da superfície (papada, lombo, barriga e pernil) de 239 carcaças processadas em uma indústria exportadora Abatedoura frigorífica de suínos (3200 animais por dia). Equipamentos que demonstraram incrementar as CT e o isolamento de Salmonella spp. nas carcaças suínas avaliadas foram submetidos a uma lista de verificação dos requisitos do design higiênico (113 itens). A superfície em contato com as carcaças das polidoras foram os chicotes, confeccionados por elastômeros, avaliados quanto à formação e remoção de biofilme em dois diferentes tipos de elastômeros e analisados quanto a sua topografia, através de Microscopia Eletrônica de Varredura (MEV). Reduções significativas (p<0,05) foram observadas nas CT de 99,85% após a escaldagem e 99,8% e 94,36%, em ambas as chamuscagens, respectivamente. Incrementos significativos (p<0,05) foram encontrados nas depilagem e em ambos os polimentos de carcaças. Na incidência de Salmonella spp. os resultados encontrados foram semelhantes; Salmonella spp. não foi detectada após a escalda e após os chamuscadores. Entretanto houve um aumento na incidência após a depiladeira e em ambas as polidoras. Estes resultados demonstraram que a depiladeira e a polidora aumentaram as contaminações microbiológicas das carcaças suínas, portanto foram diagnosticadas através de uma lista de verificação do design higiênico e demonstraram que mais de 50% dos requisitos não foram atendidos por estas máquinas, corroborando com a avaliação microbiológica. A formação e remoção de biofilme na superfície dos diferentes elastômeros, utilizados na construção dos chicotes e avaliados neste estudo, demonstraram que a maior aderência ocorreu no tempo de 180 min em ambos os materiais avaliados e que o elastômero termoplástico poliéster (ETP) demonstrou aderir menos em relação ao confeccionado de borracha natural vulcanizada. Em relação à remoção de biofilme, ambos os tempos de contato com o ácido peracético a 1% testados (1 e 10 min) demonstraram ser eficientes na remoção ou eliminação de Salmonella spp. dos corpos de prova. Na avaliação da topografia da superfície dos diferentes elastômeros, através da MEV, a borracha natural vulcanizada demonstrou possuir uma rugosidade mais elevada quando comparada ao ETP, corroborando com os resultados encontrados na aderência. / The aim of this study was to evaluate the microbial contamination, the incidence of Salmonella spp. at various stages of the pork slaughterlines and aspects of hygienic design from equipments which contributed with the pork carcass contamination. To evaluate the microbial contamination and the incidence of Salmonella spp. at various stages of the pork slaughterlines, the total counts (TC) and the incidence of Salmonella spp. were estimated in four surface locations (Ham, belly, loin and neck) on 239 pork carcasses processed through a big exporter pig-slaughtering plant (3200 animals per day). In order to evaluate important aspects of design hygienic from equipments that contributed with pork carcass contamination, the hygienic design requirements list verification (113 items) were used to do a diagnosis, to verify the biofilm formation and removal in different elastomers used in polishing food area and to compare both surface in scanning electron microscope. A significant (p <0.05) increase was found in dehairing and in both polishing stages. Similar results to Salmonella incidence were observed; the incidence increased to 16.67% from 0% on dehairing and then to 20% after the first polishing and increased 10.5% from 0% after the second polishing. When dehairing and polishing machines were diagnosed with hygienic design requirements list verification both demonstrated not attended more than 50 % of the hygienic design requirements, both machines shown poor welding, dead spaces where dirt can be retained and resulting in food contamination, for example. The biofilm formation and removal demonstrated significant difference between the time of contact with the bacterial solution (1, 30, 60 and 180 min) and the time of 180 min demonstrated greater adherence when compared to other contact time. Elastomer Termoplastic Polyester (ETP) demonstrated less adherence when compared with natural volcanic rubber. In the biofilm removals were used 180 min in contact with the bacterial suspension and 1 and 10 min in contact with the peracetic acid 1% were not observed microbial growth in both elastomers. When were compared both surface in scanning electron microscope, natural volcanic rubber demonstrated more roughness when compared with ETP. Read more
|
3 |
Avaliação do design higiênico de equipamentos que contribuem para a contaminação de carcaças na 1ª etapa de abate de suínos / Evaluation of hygienic design of equipments which contribute to pork cascasses contamination in a slaughterhouseMariot, Roberta Fogliatto January 2010 (has links)
O objetivo deste trabalho foi avaliar a contaminação microbiológica e o isolamento de Salmonella spp. em diferentes estágios da 1° etapa de um abatedouro frigorífico, além de aspectos importantes do design higiênico de equipamentos que demonstraram contribuir para a contaminação das carcaças. Para a avaliação microbiológica das carcaças suínas foram estimadas as contagens de microrganismos heterotróficos totais (CT) e a incidência de Salmonella spp. em quatro locais da superfície (papada, lombo, barriga e pernil) de 239 carcaças processadas em uma indústria exportadora Abatedoura frigorífica de suínos (3200 animais por dia). Equipamentos que demonstraram incrementar as CT e o isolamento de Salmonella spp. nas carcaças suínas avaliadas foram submetidos a uma lista de verificação dos requisitos do design higiênico (113 itens). A superfície em contato com as carcaças das polidoras foram os chicotes, confeccionados por elastômeros, avaliados quanto à formação e remoção de biofilme em dois diferentes tipos de elastômeros e analisados quanto a sua topografia, através de Microscopia Eletrônica de Varredura (MEV). Reduções significativas (p<0,05) foram observadas nas CT de 99,85% após a escaldagem e 99,8% e 94,36%, em ambas as chamuscagens, respectivamente. Incrementos significativos (p<0,05) foram encontrados nas depilagem e em ambos os polimentos de carcaças. Na incidência de Salmonella spp. os resultados encontrados foram semelhantes; Salmonella spp. não foi detectada após a escalda e após os chamuscadores. Entretanto houve um aumento na incidência após a depiladeira e em ambas as polidoras. Estes resultados demonstraram que a depiladeira e a polidora aumentaram as contaminações microbiológicas das carcaças suínas, portanto foram diagnosticadas através de uma lista de verificação do design higiênico e demonstraram que mais de 50% dos requisitos não foram atendidos por estas máquinas, corroborando com a avaliação microbiológica. A formação e remoção de biofilme na superfície dos diferentes elastômeros, utilizados na construção dos chicotes e avaliados neste estudo, demonstraram que a maior aderência ocorreu no tempo de 180 min em ambos os materiais avaliados e que o elastômero termoplástico poliéster (ETP) demonstrou aderir menos em relação ao confeccionado de borracha natural vulcanizada. Em relação à remoção de biofilme, ambos os tempos de contato com o ácido peracético a 1% testados (1 e 10 min) demonstraram ser eficientes na remoção ou eliminação de Salmonella spp. dos corpos de prova. Na avaliação da topografia da superfície dos diferentes elastômeros, através da MEV, a borracha natural vulcanizada demonstrou possuir uma rugosidade mais elevada quando comparada ao ETP, corroborando com os resultados encontrados na aderência. / The aim of this study was to evaluate the microbial contamination, the incidence of Salmonella spp. at various stages of the pork slaughterlines and aspects of hygienic design from equipments which contributed with the pork carcass contamination. To evaluate the microbial contamination and the incidence of Salmonella spp. at various stages of the pork slaughterlines, the total counts (TC) and the incidence of Salmonella spp. were estimated in four surface locations (Ham, belly, loin and neck) on 239 pork carcasses processed through a big exporter pig-slaughtering plant (3200 animals per day). In order to evaluate important aspects of design hygienic from equipments that contributed with pork carcass contamination, the hygienic design requirements list verification (113 items) were used to do a diagnosis, to verify the biofilm formation and removal in different elastomers used in polishing food area and to compare both surface in scanning electron microscope. A significant (p <0.05) increase was found in dehairing and in both polishing stages. Similar results to Salmonella incidence were observed; the incidence increased to 16.67% from 0% on dehairing and then to 20% after the first polishing and increased 10.5% from 0% after the second polishing. When dehairing and polishing machines were diagnosed with hygienic design requirements list verification both demonstrated not attended more than 50 % of the hygienic design requirements, both machines shown poor welding, dead spaces where dirt can be retained and resulting in food contamination, for example. The biofilm formation and removal demonstrated significant difference between the time of contact with the bacterial solution (1, 30, 60 and 180 min) and the time of 180 min demonstrated greater adherence when compared to other contact time. Elastomer Termoplastic Polyester (ETP) demonstrated less adherence when compared with natural volcanic rubber. In the biofilm removals were used 180 min in contact with the bacterial suspension and 1 and 10 min in contact with the peracetic acid 1% were not observed microbial growth in both elastomers. When were compared both surface in scanning electron microscope, natural volcanic rubber demonstrated more roughness when compared with ETP. Read more
|
4 |
Avaliação do design higiênico de equipamentos que contribuem para a contaminação de carcaças na 1ª etapa de abate de suínos / Evaluation of hygienic design of equipments which contribute to pork cascasses contamination in a slaughterhouseMariot, Roberta Fogliatto January 2010 (has links)
O objetivo deste trabalho foi avaliar a contaminação microbiológica e o isolamento de Salmonella spp. em diferentes estágios da 1° etapa de um abatedouro frigorífico, além de aspectos importantes do design higiênico de equipamentos que demonstraram contribuir para a contaminação das carcaças. Para a avaliação microbiológica das carcaças suínas foram estimadas as contagens de microrganismos heterotróficos totais (CT) e a incidência de Salmonella spp. em quatro locais da superfície (papada, lombo, barriga e pernil) de 239 carcaças processadas em uma indústria exportadora Abatedoura frigorífica de suínos (3200 animais por dia). Equipamentos que demonstraram incrementar as CT e o isolamento de Salmonella spp. nas carcaças suínas avaliadas foram submetidos a uma lista de verificação dos requisitos do design higiênico (113 itens). A superfície em contato com as carcaças das polidoras foram os chicotes, confeccionados por elastômeros, avaliados quanto à formação e remoção de biofilme em dois diferentes tipos de elastômeros e analisados quanto a sua topografia, através de Microscopia Eletrônica de Varredura (MEV). Reduções significativas (p<0,05) foram observadas nas CT de 99,85% após a escaldagem e 99,8% e 94,36%, em ambas as chamuscagens, respectivamente. Incrementos significativos (p<0,05) foram encontrados nas depilagem e em ambos os polimentos de carcaças. Na incidência de Salmonella spp. os resultados encontrados foram semelhantes; Salmonella spp. não foi detectada após a escalda e após os chamuscadores. Entretanto houve um aumento na incidência após a depiladeira e em ambas as polidoras. Estes resultados demonstraram que a depiladeira e a polidora aumentaram as contaminações microbiológicas das carcaças suínas, portanto foram diagnosticadas através de uma lista de verificação do design higiênico e demonstraram que mais de 50% dos requisitos não foram atendidos por estas máquinas, corroborando com a avaliação microbiológica. A formação e remoção de biofilme na superfície dos diferentes elastômeros, utilizados na construção dos chicotes e avaliados neste estudo, demonstraram que a maior aderência ocorreu no tempo de 180 min em ambos os materiais avaliados e que o elastômero termoplástico poliéster (ETP) demonstrou aderir menos em relação ao confeccionado de borracha natural vulcanizada. Em relação à remoção de biofilme, ambos os tempos de contato com o ácido peracético a 1% testados (1 e 10 min) demonstraram ser eficientes na remoção ou eliminação de Salmonella spp. dos corpos de prova. Na avaliação da topografia da superfície dos diferentes elastômeros, através da MEV, a borracha natural vulcanizada demonstrou possuir uma rugosidade mais elevada quando comparada ao ETP, corroborando com os resultados encontrados na aderência. / The aim of this study was to evaluate the microbial contamination, the incidence of Salmonella spp. at various stages of the pork slaughterlines and aspects of hygienic design from equipments which contributed with the pork carcass contamination. To evaluate the microbial contamination and the incidence of Salmonella spp. at various stages of the pork slaughterlines, the total counts (TC) and the incidence of Salmonella spp. were estimated in four surface locations (Ham, belly, loin and neck) on 239 pork carcasses processed through a big exporter pig-slaughtering plant (3200 animals per day). In order to evaluate important aspects of design hygienic from equipments that contributed with pork carcass contamination, the hygienic design requirements list verification (113 items) were used to do a diagnosis, to verify the biofilm formation and removal in different elastomers used in polishing food area and to compare both surface in scanning electron microscope. A significant (p <0.05) increase was found in dehairing and in both polishing stages. Similar results to Salmonella incidence were observed; the incidence increased to 16.67% from 0% on dehairing and then to 20% after the first polishing and increased 10.5% from 0% after the second polishing. When dehairing and polishing machines were diagnosed with hygienic design requirements list verification both demonstrated not attended more than 50 % of the hygienic design requirements, both machines shown poor welding, dead spaces where dirt can be retained and resulting in food contamination, for example. The biofilm formation and removal demonstrated significant difference between the time of contact with the bacterial solution (1, 30, 60 and 180 min) and the time of 180 min demonstrated greater adherence when compared to other contact time. Elastomer Termoplastic Polyester (ETP) demonstrated less adherence when compared with natural volcanic rubber. In the biofilm removals were used 180 min in contact with the bacterial suspension and 1 and 10 min in contact with the peracetic acid 1% were not observed microbial growth in both elastomers. When were compared both surface in scanning electron microscope, natural volcanic rubber demonstrated more roughness when compared with ETP. Read more
|
5 |
Reanimador manual: quando trocar no mesmo pacienteGomes, Giselle Pinheiro Lima Aires 31 March 2016 (has links)
Submitted by JÚLIO HEBER SILVA (julioheber@yahoo.com.br) on 2016-11-09T14:03:31Z
No. of bitstreams: 2
Tese - Giselle Pinheiro Lima Aires Gomes - 2016.pdf: 3158182 bytes, checksum: ba9d36d200045df31c0cbd0972cc94a2 (MD5)
license_rdf: 0 bytes, checksum: d41d8cd98f00b204e9800998ecf8427e (MD5) / Approved for entry into archive by Jaqueline Silva (jtas29@gmail.com) on 2016-11-10T17:49:41Z (GMT) No. of bitstreams: 2
Tese - Giselle Pinheiro Lima Aires Gomes - 2016.pdf: 3158182 bytes, checksum: ba9d36d200045df31c0cbd0972cc94a2 (MD5)
license_rdf: 0 bytes, checksum: d41d8cd98f00b204e9800998ecf8427e (MD5) / Made available in DSpace on 2016-11-10T17:49:41Z (GMT). No. of bitstreams: 2
Tese - Giselle Pinheiro Lima Aires Gomes - 2016.pdf: 3158182 bytes, checksum: ba9d36d200045df31c0cbd0972cc94a2 (MD5)
license_rdf: 0 bytes, checksum: d41d8cd98f00b204e9800998ecf8427e (MD5)
Previous issue date: 2016-03-31 / Fundação de Amparo à Pesquisa do Estado de Goiás - FAPEG / INTRODUCTION: The manual resuscitator is a widely used respiratory assist device
that has been reported to be a reservoir and a source of contamination from various
microorganisms. At present, there is no criteria to replacement of manual resuscitator
when it is in successive use in the same patient. AIM: To evaluate the safest amount
of time the manual resuscitator can be successively used in the same patient.
METHODS: An open, prospective cohort study was conducted from October to
November, 2014 using 30 patient connector valves from manual resuscitator devices
obtained from Intensive Care Units of a general hospital located in a region north of
Brazil. The samples were collected through swab friction on the manual resuscitator
that was used by the same patient, at zero (ready use), 24 and 48 hours. Bacterial
identification and antibiotic susceptibility were performed automatically (Vitek 2
Compact®). RESULTS: Of the 30 resuscitators evaluated, 20 (66.6%) were found to
be contaminated. There was a significant difference between the microbial load on
the manual resuscitators in use at zero and 24 hours (p = 0.03). Associated risk
factors for the contamination of manual resuscitators identified were frequency and
time of use. The presence of visible soil was not detected on 19 manual resuscitators
in use, however, 95.0% were contaminated. The number of microorganisms isolated
at zero, 24 and 48 hours were five, 11 and 24, respectively. Thirteen devices were
contaminated with two or more bacterial species. Of the Gram-positive cocci, 38.9%
(n = 18) were methicillin-resistant Staphylococcus aureus and 11.1% were
methicillin-resistant coagulase-negative Staphylococcus, all were constitutive MLSB
resistant. Of the Gram-negative rods (n = 36), Acinetobacter baumannii (36.1%),
Pseudomonas aeruginosa (19.4%), Serratia marcescens (22.2%) and Proteus spp.
(8.3%) dominated. Over 50% of these were resistant to carbapenems, second, third
and fourth cephalosporins generations, and ampicillin/sulbactam. CONCLUSION:
Manual resuscitators in successive use in the same patient were contaminated even
in the absence of visible dirt. Multi- and extensively-resistant bacteria of clinical
importance were also detected. Frequency and time of use were identified as risk
factors in the contamination of manual resuscitators. The longer the time of use, the
greater the number of contaminated resuscitators and bacterial species isolated.
These results point to failures in reprocessing, and therefore highlights the
importance of having thorough discussions among regulators about the
recommendations for the reprocessing of semi-critical medical devices, especially,
those for ventilatory assistance. Furthermore, the results highlight the need to
replace manual resuscitators every 24 hours after use as a strategy for infection
control and to minimize the risk of re-colonization or -infection of the respiratory tract. / INTRODUÇÃO: O reanimador manual é um dispositivo de assistência respiratória
amplamente utilizado que tem sido reportado como reservatório e fonte de
contaminação por diversos micro-organismos, e ainda não apresenta critérios
definidos para a troca quando em uso sucessivo no mesmo paciente. OBJETIVO:
Avaliar o tempo de uso seguro do reanimador manual em uso sucessivo no mesmo
paciente. MÉTODO: Trata-se de uma coorte aberta prospectiva, realizada de
outubro a novembro de 2014 em 30 válvulas do reanimador manual (conector do
paciente) em Unidades de Tratamento Intensivo de um hospital geral da região
Norte do Brasil. As amostras foram coletadas por meio de fricção de swab em
reanimador manual em uso no mesmo paciente nos tempos zero (pronto uso), 24 e
48 horas. A identificação bacteriana e o antibiograma foram automatizados (Vitek 2
Compact®). RESULTADOS: Dos 30 reanimadores avaliados, 20 (66,6%) estavam
contaminados. A carga microbiana entre os tempos zero e 24h de uso dos
reanimadores manuais apresentou diferença estatística significativa (p=0,03). A
frequência e o tempo de uso foram identificados como fatores de risco associados
para a contaminação de reanimadores manuais em uso. Dos 19 reanimadores
manuais avaliados como visivelmente limpos, 95,0% apresentaram contaminados.
Foram isolados cinco, 11 e 24 tipos de micro-organismos nos tempos zero, 24 e 48
horas respectivamente. Treze dispositivos estavam contaminados por duas ou mais
espécies bacterianas. Dentre os cocos gram-positivos (n=18), 38,9% eram
Staphylococcus aureus resistentes à meticilina e 11,1% Staphylococcus coagulase
negativos resistentes à meticilina, todos com resistência constitutiva ao grupo MLSB.
Dentre os bastonetes gram-negativos (n=36), predominaram Acinetobacter
baumannii (36,1%), Pseudomonas aeruginosa (19,4%), Serratia marcescens
(22,2%) e Proteus spp. (8,3%). Mais de 50% destes apresentaram resistência aos
carbapenens, cefalosporinas de segunda, terceira e quarta gerações, e
ampicilina/sulbactam. CONCLUSÃO: Os reanimadores manuais em uso sucessivo
no mesmo paciente estavam contaminados, mesmo na ausência de sujidade visível,
sendo o tempo e a frequência de uso do dispositivo fatores de risco identificados.
Bactérias multirresistentes e extensivamente resistentes de importância clínica foram
detectadas. Quanto maior o tempo de uso, maior o número de reanimadores
contaminados e de espécies bacterianas isoladas. Os resultados apontam para
falhas no processamento, e indicam necessidade de discussão entre os órgãos
regulamentadores sobre a recomendação para o processamento de produtos para a
saúde semicríticos, em especial, os de assistência ventilatória e demonstram ainda a
necessidade de troca do reanimador manual a cada 24 horas de uso, como
estratégia primária para minimizar o risco de (re)colonização que poderá resultar em
infecção do trato respiratório. Read more
|