• Refine Query
  • Source
  • Publication year
  • to
  • Language
  • 2
  • 2
  • 1
  • Tagged with
  • 5
  • 4
  • 4
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • 2
  • About
  • The Global ETD Search service is a free service for researchers to find electronic theses and dissertations. This service is provided by the Networked Digital Library of Theses and Dissertations.
    Our metadata is collected from universities around the world. If you manage a university/consortium/country archive and want to be added, details can be found on the NDLTD website.
1

Aktivierungsexperimente mit Neutronen mittlerer Energien und die Produktion kosmogener Nuklide in extraterrestrischer Materie

Neumann, Sonja. January 1999 (has links) (PDF)
Hannover, Universiẗat, Diss., 1999.
2

Neutron activation as an independent indicator of expected total yield in the production of 82Sr and 68Ge with 66 MeV protons

Vermeulen, C., Steyn, G. F., van der Meulen, N. P. 19 May 2015 (has links) (PDF)
Introduction A method based on neutron activation is being developed to assist in resolving discrepancies between the expected yield and actual yield of radionuclides produced with the vertical-beam target station (VBTS) at iThemba LABS. The VBTS is routinely employed for multi-Ci batch productions of the radionuclide pairs 22Na/68Ga and 82Sr/68Ga using standardized natMg/natGa and natRb/natGa tandem targets, respectively [1]. The metal-clad target discs are bombarded with a primary beam of 66 MeV protons at an intensity of nominally 250 µA. The encapsulation materials are either Nb (for Mg and Ga) or stainless steel (for Rb) which serve to contain the molten target materials during bombardment and act as a barrier to the high-velocity cooling water which surrounds the targets in a 4π geometry. The natRb/natGa targets are typically bombarded according to a two-week cycle while natMg/natGa targets are bombarded on an ad-hoc basis, depending on a somewhat unpredictable 22Na demand. A too-large deviation between expected yield and actual yield has at times plagued this programme. These deviations can manifest both as an apparent loss or an apparent gain (relative to the expected yield) by up to about 15% in either direction. The resulting uncertainty of up to 30% (in the worst case) from one production batch to another can be costly and is unacceptable in a large-scale production regimen. This phenomenon is believed to be brought about by two types of problems: (1) Production losses, e.g. during the radio-chemical separation process or incomplete recovery of activated target material during the decapsulation step. (2) Incorrect values obtained for the accumulated proton charge. A problem of type (1) will always result in a loss of yield. A problem of type (2) can manifest as an apparent loss or gain. In an effort to get a handle on this second type of problem, neutron activation of suitable material samples, embedded in a target holder, is being investigated as an independent indicator of the total yield. For this purpose, samples of Co, Mn, Ni and Zn were activated during production runs and Co was found to be the most appropriate. Preliminary results will be presented after first discussing why the determination of the accumulated pro-ton charge is a problem with the VBTS. Materials and Methods The VBTS consists of a central region in which a target holder is located during bombardment as well as two half-cylindrical radiation shields which completely surround the target. The shields can be moved away from the central region on dedicated rails, e.g. when repairs or maintenance is required. FIGURE 1 shows the VBTS with the shields moved to the “open” position. As some components of the station are located below the vault floor, with the target position near floor level, it proved difficult to electrically isolate the VBTS as was done for the two horizontal-beam target stations at iThemba LABS [1]. The VBTS does not act as a Faraday cup like the other target stations. Instead, the beam current and accumulated charge is measured by means of a calibrated capacitive probe [1,2]. There appears to be a variation in the response of the capacitive probe, sensitive to the beam microstructure, in particular a dependence on the beam packet length. This problem is not yet fully resolved. FIGURE 2 (a) shows the beamstop of a VBTS target holder with several Co samples mounted on the outside as well as one each of Ni, Mn and Zn. The samples are small “tablets” with a 10 mm diameter and 1 mm thickness. The reactions of interest are 59Co(n,γ)60Co, 59Co(n,3n)57Co, nat-Ni(n,X)60Co, natNi(n,X)57Co, natZn(n,X)65Zn and 55Mn(n,2n)54Mn. The relevant half-lives are 60Co(5.271 a), 57Co(271.8 d), 65Zn(244.3 d) and 54Mn(312.2 d). The half-life should be long compared to the two-week cycle in order to reduce the dependence on the exact beam history, which is very fragmented over any production period. In this respect, 60Co is considered to be particularly attractive as its long half-life of more than 5 years leads to a negligible effect by the beam history. Note that the tandem targets, shown in FIGURE 2 (b), are mounted just upstream of the beamstop – in fact, the targets and beamstop form a single unit before being fitted into the target holder. At the end of bombardment, all samples were assayed for their characteristic γ-emissions using standard off-line γ-ray spectrometry with an HPGe detector connected to a Genie 2000 MCA. Calculations of the neutron fluence density in the central sample volume on the beamstop were also performed using the Monte Carlo radiation transport code MCNPX. For these calculations, the entire VBTS, a Rb/Ga target and the vault walls were included in the model. Results and Conclusion All samples activated significantly – copious amounts of 60Co were detected in the Co discs after a two-week run. The neutron fluence density for the case of a 250 µA, 66 MeV proton beam on a natRb/natGa tandem target is shown in FIGURE 3. The dominance of low-energy neutrons is evident, which is in part due to the large amount of paraffin-wax shielding material in close proximity to the target. While reactions such as the (n,2n) and (n,3n) would be sensitive to the more energetic part of the neutron spectrum, the (n,γ) capture reaction benefits from the large low-energy component. This explains the copious amounts of 60Co formed. It was therefore decided to only retain the central Co sample for subsequent bombardments, as shown in FIGURE 4. The first results are shown in TABLE 1. The accumulated charge as obtained from the capacitive probe (Q), the specific 60Co activity (A) at the end of bombardment (EOB), and their ratio (A/Q) are presented in the table, together with the deviation of individual ratios relative to their average for the case of the Mg/Ga tandem tar-gets only. Note that all samples were counted until the statistical uncertainties were negligible. Any systematic uncertainties are ignored at this stage as they are considered to remain the same from one batch production to another. For the sake of argument, the average value of the ratio is taken as the expected value. A positive deviation of the A/Q value is then indicative of a too-small value of the accumulated charge obtained from the capacitive probe, leading to a corresponding overproduction. Likewise, a negative value is indicative of a too-large value of the accumulated charge, leading to a corresponding underproduction. It is certainly true that the data in TABLE 1 are currently very limited. It is envisaged, however, that with time the growing database of values will assist in reducing the uncertainty in determining the accumulated charge and reduce the discrepancies between predicted and actual yields significantly. TABLE 1 illuminates the underlying problem satisfactorily. The four Mg/Ga tandem target bombardments, on identical targetry, were performed successively. The neutron activation correlates well the with actual yields, pointing directly to the current integration as the main source of error. The method already proves to be useful. An indication of an over or underprediction can be obtained prior to the target processing by recovering and measuring the Co disc. This in-formation can be used to make a decision concerning the present batch production and/or the subsequent one. One can either add beam to the present production target and/or in-crease/reduce the total beam on the subsequent production target to compensate for an expected overproduction or shortfall. In conclusion, we would like to stress that the capacitive probes show great promise and that better understanding and/or possibly some development of their signal processing algorithm may improve their ability to measure the accumulated charge to the desired accuracy. Segmented capacitive probes used at iThemba LABS and elsewhere for beam position measurement [1,3] are not affected by beam microstructure as only the ratios of the signal strengths on the different sectors are important. In this case, changes in response affect all sec-tors equally and the ratios are unaffected.
3

Neutron activation as an independent indicator of expected total yield in the production of 82Sr and 68Ge with 66 MeV protons

Vermeulen, C., Steyn, G. F., van der Meulen, N. P. January 2015 (has links)
Introduction A method based on neutron activation is being developed to assist in resolving discrepancies between the expected yield and actual yield of radionuclides produced with the vertical-beam target station (VBTS) at iThemba LABS. The VBTS is routinely employed for multi-Ci batch productions of the radionuclide pairs 22Na/68Ga and 82Sr/68Ga using standardized natMg/natGa and natRb/natGa tandem targets, respectively [1]. The metal-clad target discs are bombarded with a primary beam of 66 MeV protons at an intensity of nominally 250 µA. The encapsulation materials are either Nb (for Mg and Ga) or stainless steel (for Rb) which serve to contain the molten target materials during bombardment and act as a barrier to the high-velocity cooling water which surrounds the targets in a 4π geometry. The natRb/natGa targets are typically bombarded according to a two-week cycle while natMg/natGa targets are bombarded on an ad-hoc basis, depending on a somewhat unpredictable 22Na demand. A too-large deviation between expected yield and actual yield has at times plagued this programme. These deviations can manifest both as an apparent loss or an apparent gain (relative to the expected yield) by up to about 15% in either direction. The resulting uncertainty of up to 30% (in the worst case) from one production batch to another can be costly and is unacceptable in a large-scale production regimen. This phenomenon is believed to be brought about by two types of problems: (1) Production losses, e.g. during the radio-chemical separation process or incomplete recovery of activated target material during the decapsulation step. (2) Incorrect values obtained for the accumulated proton charge. A problem of type (1) will always result in a loss of yield. A problem of type (2) can manifest as an apparent loss or gain. In an effort to get a handle on this second type of problem, neutron activation of suitable material samples, embedded in a target holder, is being investigated as an independent indicator of the total yield. For this purpose, samples of Co, Mn, Ni and Zn were activated during production runs and Co was found to be the most appropriate. Preliminary results will be presented after first discussing why the determination of the accumulated pro-ton charge is a problem with the VBTS. Materials and Methods The VBTS consists of a central region in which a target holder is located during bombardment as well as two half-cylindrical radiation shields which completely surround the target. The shields can be moved away from the central region on dedicated rails, e.g. when repairs or maintenance is required. FIGURE 1 shows the VBTS with the shields moved to the “open” position. As some components of the station are located below the vault floor, with the target position near floor level, it proved difficult to electrically isolate the VBTS as was done for the two horizontal-beam target stations at iThemba LABS [1]. The VBTS does not act as a Faraday cup like the other target stations. Instead, the beam current and accumulated charge is measured by means of a calibrated capacitive probe [1,2]. There appears to be a variation in the response of the capacitive probe, sensitive to the beam microstructure, in particular a dependence on the beam packet length. This problem is not yet fully resolved. FIGURE 2 (a) shows the beamstop of a VBTS target holder with several Co samples mounted on the outside as well as one each of Ni, Mn and Zn. The samples are small “tablets” with a 10 mm diameter and 1 mm thickness. The reactions of interest are 59Co(n,γ)60Co, 59Co(n,3n)57Co, nat-Ni(n,X)60Co, natNi(n,X)57Co, natZn(n,X)65Zn and 55Mn(n,2n)54Mn. The relevant half-lives are 60Co(5.271 a), 57Co(271.8 d), 65Zn(244.3 d) and 54Mn(312.2 d). The half-life should be long compared to the two-week cycle in order to reduce the dependence on the exact beam history, which is very fragmented over any production period. In this respect, 60Co is considered to be particularly attractive as its long half-life of more than 5 years leads to a negligible effect by the beam history. Note that the tandem targets, shown in FIGURE 2 (b), are mounted just upstream of the beamstop – in fact, the targets and beamstop form a single unit before being fitted into the target holder. At the end of bombardment, all samples were assayed for their characteristic γ-emissions using standard off-line γ-ray spectrometry with an HPGe detector connected to a Genie 2000 MCA. Calculations of the neutron fluence density in the central sample volume on the beamstop were also performed using the Monte Carlo radiation transport code MCNPX. For these calculations, the entire VBTS, a Rb/Ga target and the vault walls were included in the model. Results and Conclusion All samples activated significantly – copious amounts of 60Co were detected in the Co discs after a two-week run. The neutron fluence density for the case of a 250 µA, 66 MeV proton beam on a natRb/natGa tandem target is shown in FIGURE 3. The dominance of low-energy neutrons is evident, which is in part due to the large amount of paraffin-wax shielding material in close proximity to the target. While reactions such as the (n,2n) and (n,3n) would be sensitive to the more energetic part of the neutron spectrum, the (n,γ) capture reaction benefits from the large low-energy component. This explains the copious amounts of 60Co formed. It was therefore decided to only retain the central Co sample for subsequent bombardments, as shown in FIGURE 4. The first results are shown in TABLE 1. The accumulated charge as obtained from the capacitive probe (Q), the specific 60Co activity (A) at the end of bombardment (EOB), and their ratio (A/Q) are presented in the table, together with the deviation of individual ratios relative to their average for the case of the Mg/Ga tandem tar-gets only. Note that all samples were counted until the statistical uncertainties were negligible. Any systematic uncertainties are ignored at this stage as they are considered to remain the same from one batch production to another. For the sake of argument, the average value of the ratio is taken as the expected value. A positive deviation of the A/Q value is then indicative of a too-small value of the accumulated charge obtained from the capacitive probe, leading to a corresponding overproduction. Likewise, a negative value is indicative of a too-large value of the accumulated charge, leading to a corresponding underproduction. It is certainly true that the data in TABLE 1 are currently very limited. It is envisaged, however, that with time the growing database of values will assist in reducing the uncertainty in determining the accumulated charge and reduce the discrepancies between predicted and actual yields significantly. TABLE 1 illuminates the underlying problem satisfactorily. The four Mg/Ga tandem target bombardments, on identical targetry, were performed successively. The neutron activation correlates well the with actual yields, pointing directly to the current integration as the main source of error. The method already proves to be useful. An indication of an over or underprediction can be obtained prior to the target processing by recovering and measuring the Co disc. This in-formation can be used to make a decision concerning the present batch production and/or the subsequent one. One can either add beam to the present production target and/or in-crease/reduce the total beam on the subsequent production target to compensate for an expected overproduction or shortfall. In conclusion, we would like to stress that the capacitive probes show great promise and that better understanding and/or possibly some development of their signal processing algorithm may improve their ability to measure the accumulated charge to the desired accuracy. Segmented capacitive probes used at iThemba LABS and elsewhere for beam position measurement [1,3] are not affected by beam microstructure as only the ratios of the signal strengths on the different sectors are important. In this case, changes in response affect all sec-tors equally and the ratios are unaffected.
4

Untersuchungen zur Sauerstoffausscheidung in hoch bordotiertem Silicium

Zschorsch, Markus 24 July 2009 (has links) (PDF)
Die Sauerstoffausscheidung in Silicium wird durch die Höhe der Bordotierung beeinflusst. Mit dem Ziel der Aufklärung der Mechanismen bei der Ausscheidung wurden verschiedene physikalische Messmethoden angepasst und ein breiter Borkonzentrationsbereich charakterisiert. Es wurden die frühen Phasen der Ausscheidungsbildung sowie Komplexbildung als auch das Wachstum der Sauerstoffausscheidungen untersucht. Mithilfe einer neuen Methodenkombination aus alkalischem Ätzen und FTIR-Spektroskopie konnten verschiedene Bor-Komplexe nachgewiesen werden. Die Erkenntnisse über deren Existenz sowie Kinetik wurden numerisch umgesetzt. Mittels Kleinwinkel-Neutronenstreuung wurden erstmals in den frühen Phasen der Sauerstoffausscheidung deren Form und Größe in Abhängigkeit der Borkonzentration bestimmt. Die physikalischen Prozesse, die zu einem beschleunigten als auch anomalen Ausscheidungsverhalten in Abhängigkeit der Borkonzentration führen, konnten teilweise aufgeklärt werden.
5

Untersuchungen zur Sauerstoffausscheidung in hoch bordotiertem Silicium

Zschorsch, Markus 14 December 2007 (has links)
Die Sauerstoffausscheidung in Silicium wird durch die Höhe der Bordotierung beeinflusst. Mit dem Ziel der Aufklärung der Mechanismen bei der Ausscheidung wurden verschiedene physikalische Messmethoden angepasst und ein breiter Borkonzentrationsbereich charakterisiert. Es wurden die frühen Phasen der Ausscheidungsbildung sowie Komplexbildung als auch das Wachstum der Sauerstoffausscheidungen untersucht. Mithilfe einer neuen Methodenkombination aus alkalischem Ätzen und FTIR-Spektroskopie konnten verschiedene Bor-Komplexe nachgewiesen werden. Die Erkenntnisse über deren Existenz sowie Kinetik wurden numerisch umgesetzt. Mittels Kleinwinkel-Neutronenstreuung wurden erstmals in den frühen Phasen der Sauerstoffausscheidung deren Form und Größe in Abhängigkeit der Borkonzentration bestimmt. Die physikalischen Prozesse, die zu einem beschleunigten als auch anomalen Ausscheidungsverhalten in Abhängigkeit der Borkonzentration führen, konnten teilweise aufgeklärt werden.

Page generated in 0.089 seconds