West Virginia University
The effects of cortical bone viscoelasticity on the fixation/stability of cemented and cementless femoral implants: A finite element analysis.
Abstract
dc:description.abstractThere are two classifications of femoral implants ñ cemented and cementless. The object of this study was to assess the effect of bone viscoelasticity on cemented and cementless implant stability. A cemented implant design was evaluated via a 3-D FE model. The results showed a small 1.41% increase in stem subsidence with inclusion of bone viscoelasticity. This is explained by the low stresses at the cement-bone interface. A 2-D FE model was utilized to evaluate the effects of coefficient of friction, press-fit and bone viscoelasticity on stem fixation. The push-out load increased with µ for each value of δ, but minimal increases in the push-out load (2.9% to 4.9%) were observed for δ > 0.10mm. Bone viscoelasticity reduced the push-out load (2.6% to 82.6%) in all cases. It is concluded that there is a ìthresholdî with respect to the degree of press-fit beyond which no additional gains in fixation are achieved.
Degree
thesis:*- Name thesis:degree_name
- MS
- Level thesis:degree_level
- Thesis
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Shultz, Travis Ray
- Contributors dc:contributor
-
- Timothy L. Norman
Identifiers
dc:identifier.*- Identifier
- https://researchrepository.wvu.edu/etd/10665