Back to search

West Virginia University

Comparative Analysis of Electrode Encapsulation Reliability: Planar vs. Thermoformed Interdigitated Neural Electrodes in Biomedical Engineering

Abstract

dc:description.abstract

Durable dielectric encapsulation is crucial for the long-term function of implantable neural electrodes; however, the reliability of curved, thermoformed coatings remains underexplored. This study addresses that gap by comparing planar (non-thermoformed) and thermoformed interdigitated neural electrodes (IDEs) encapsulated in Polyimide-2611. Both device types underwent accelerated in vitro aging in phosphate‑buffered saline (PBS) at 67 °C for approximately 91 days, simulating two years of physiological exposure. Post-accelerated aging evaluation included electrochemical impedance spectroscopy (EIS) from 0.1 Hz to 100 kHz to monitor moisture‑induced leakage, and high‑voltage breakdown testing under saline immersion to assess dielectric strength. After aging, all electrodes maintained impedance magnitudes above 10⁹ Ω at 0.1 Hz, 10⁶ Ω at 1 kHz, and 10⁴ Ω at 100 kHz. Initial breakdown voltages for Channel 1 averaged 3.4 kV for planar electrodes and 3.2 kV for thermoformed electrodes, indicating comparable insulation performance. Failure‑mode analysis localized leaks to the integration procedure at the solder pad interface rather than to the bulk polyimide film. These results demonstrate that thermoforming enables anatomically conformal electrode geometries without compromising long‑term encapsulation integrity, supporting the development of flexible neural interfaces.

Degree

thesis:*
Name thesis:degree_name
MS
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Chemical and Biomedical Engineering

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Peyerl, Tatum
Contributors dc:contributor
  • Stephen Cain
  • Loren Rieth
  • Xueyan Song

Subjects

dc:subject × 10

Identifiers

dc:identifier.*
Identifier
<p><a href="https://orcid.org/0009-0009-8310-7244">https://orcid.org/0009-0009-8310-7244</a></p>
https://researchrepository.wvu.edu/etd/13011

Chain of custody

source
Harvested from
West Virginia University
Base URL
researchrepository.wvu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Peyerl, Tatum. Comparative Analysis of Electrode Encapsulation Reliability: Planar vs. Thermoformed Interdigitated Neural Electrodes in Biomedical Engineering. Thesis thesis, https://doi.org/10.33915/etd.13011