Spelling suggestions: "subject:"mittelohrchirurgie"" "subject:"mittelohrchirurgische""
1 |
Die Genauigkeit der menschlichen Hand im Vergleich mit einem Mikromanipulator- präklinische Evaluation für die OhrchirurgieRunge, Annette 06 August 2012 (has links) (PDF)
Manual accuracy in microsurgery is reduced by tremor and limited access. A surgical approach through the middle ear also puts delicate structures at risk, while the surgeon is often working at an unergonomic position. At this point a micromanipulator could have a positive influence. A system was developed to measure “working accuracy”, time and precision during manipulation in the middle ear. 10 ENT- surgeons simulated a perforation of the stapedial footplate on a modified 3D print of a human skull in a mock OR. Each trial was repeated more than 200 times aiming manually and using a micro-manipulator. Data of over 4000 measurements was tested and graphically processed. Work strain was evaluated
with a questionnaire. Accuracy for manual and micromanipulator perforation revealed a small
difference. Learning curves showed a stronger decrease both in deviation and time when the micromanipulator was used. Also a lower work strain was apparent. The micromanipulator has the potential as an aiding device in ear surgery. / Die manuelle Genauigkeit in der Mikrochirurgie wird duch Tremor und limitierten Zugang eingeschränkt. Ein chirurgischer Eingriff am Mittelohr birgt außerdem ein Verletzungspotential für empfindliche anatomische Strukturen. Überdies ist die Sitzposition des Operateurs oft unergonomisch. Ein neuartiger Mikromanipulator kann auf diese Faktoren einen positiven Einfluss haben. Eine spezielle Software wurde entwickelt und Genauigkeit, Zeit und Präzision bei einem Eingriff am Mittelohr zu bestimmen. 10 Kopf- Hals- Chirurgen simulierten die Perforation der Steigbügelfußplatte an einem 3D Modell eines menschlichen Schädels in einem Demonstrations- OP. Jeder Versuch wurde mehr als 200 mal zunächst manuell und später mit Hilfe des Mikromanipulators wiederholt. Die Daten von mehr als 4000 Messversuchen wurden getestet und grafisch dargestellt. Die Arbeitsbelastung wure mittels eines Fragebogens evaluiert.
Manuelle und mikromanipulatorgestütze Genauigkeit zeigten einen signifikanten, jedoch ,absolut betrachtet,sehr geringen Unterschied der Genauigkeit. Die Lernkurven zeigten einen steileren Verlauf sowohl im Hinblick auf Genauigkeit als auch Versuchszeit, wenn der Mikromanipulator zur Anwendung kam. Weiterhin war eine geringere Arbeitsbelastung zu erkennen. Der Mikromanipulator birgt Potential als kompaktes Hilfsmittel für die Ohrchirurgie.
|
2 |
Die Genauigkeit der menschlichen Hand im Vergleich mit einem Mikromanipulator- präklinische Evaluation für die OhrchirurgieRunge, Annette 29 May 2012 (has links)
Manual accuracy in microsurgery is reduced by tremor and limited access. A surgical approach through the middle ear also puts delicate structures at risk, while the surgeon is often working at an unergonomic position. At this point a micromanipulator could have a positive influence. A system was developed to measure “working accuracy”, time and precision during manipulation in the middle ear. 10 ENT- surgeons simulated a perforation of the stapedial footplate on a modified 3D print of a human skull in a mock OR. Each trial was repeated more than 200 times aiming manually and using a micro-manipulator. Data of over 4000 measurements was tested and graphically processed. Work strain was evaluated
with a questionnaire. Accuracy for manual and micromanipulator perforation revealed a small
difference. Learning curves showed a stronger decrease both in deviation and time when the micromanipulator was used. Also a lower work strain was apparent. The micromanipulator has the potential as an aiding device in ear surgery.:Inhaltsverzeichnis 2
Bibliografische Beschreibung 3
Referat 3
1 Einführung 4
1.1. Mittelohrchirurgie= Mikrochirurgie 4
1.1.1. Stapedotomie- Operationsprinzip 4
1.1.2. Mögliche Komplikationen einer Stapedotomie durch manuelle Manipulation 4
1.2. Chirurgische Genauigkeit 5
1.3. Störfaktoren der Genauigkeit bei einem mittelohrchirurgischen Eingriff 6
1.3.1. Physiologische Limitationen der menschlichen Hand 6
1.3.2. Ergonomie des mittelohrchirurgischen Eingriffs 7
1.3.3. Besondere Bedingungen des mikrochirurgischen Eingriffs 8
1.4. Chirurgische Assistenzsysteme 8
1.5. Ziel der Arbeit 13
2 Publikationsmanuskript 15
3 Zusammenfassung 24
4 Literaturverzeichnis 30
Erklärung über die eigenständige Abfassung der Arbeit 36
Curriculum Vitae 37
Danksagung 39 / Die manuelle Genauigkeit in der Mikrochirurgie wird duch Tremor und limitierten Zugang eingeschränkt. Ein chirurgischer Eingriff am Mittelohr birgt außerdem ein Verletzungspotential für empfindliche anatomische Strukturen. Überdies ist die Sitzposition des Operateurs oft unergonomisch. Ein neuartiger Mikromanipulator kann auf diese Faktoren einen positiven Einfluss haben. Eine spezielle Software wurde entwickelt und Genauigkeit, Zeit und Präzision bei einem Eingriff am Mittelohr zu bestimmen. 10 Kopf- Hals- Chirurgen simulierten die Perforation der Steigbügelfußplatte an einem 3D Modell eines menschlichen Schädels in einem Demonstrations- OP. Jeder Versuch wurde mehr als 200 mal zunächst manuell und später mit Hilfe des Mikromanipulators wiederholt. Die Daten von mehr als 4000 Messversuchen wurden getestet und grafisch dargestellt. Die Arbeitsbelastung wure mittels eines Fragebogens evaluiert.
Manuelle und mikromanipulatorgestütze Genauigkeit zeigten einen signifikanten, jedoch ,absolut betrachtet,sehr geringen Unterschied der Genauigkeit. Die Lernkurven zeigten einen steileren Verlauf sowohl im Hinblick auf Genauigkeit als auch Versuchszeit, wenn der Mikromanipulator zur Anwendung kam. Weiterhin war eine geringere Arbeitsbelastung zu erkennen. Der Mikromanipulator birgt Potential als kompaktes Hilfsmittel für die Ohrchirurgie.:Inhaltsverzeichnis 2
Bibliografische Beschreibung 3
Referat 3
1 Einführung 4
1.1. Mittelohrchirurgie= Mikrochirurgie 4
1.1.1. Stapedotomie- Operationsprinzip 4
1.1.2. Mögliche Komplikationen einer Stapedotomie durch manuelle Manipulation 4
1.2. Chirurgische Genauigkeit 5
1.3. Störfaktoren der Genauigkeit bei einem mittelohrchirurgischen Eingriff 6
1.3.1. Physiologische Limitationen der menschlichen Hand 6
1.3.2. Ergonomie des mittelohrchirurgischen Eingriffs 7
1.3.3. Besondere Bedingungen des mikrochirurgischen Eingriffs 8
1.4. Chirurgische Assistenzsysteme 8
1.5. Ziel der Arbeit 13
2 Publikationsmanuskript 15
3 Zusammenfassung 24
4 Literaturverzeichnis 30
Erklärung über die eigenständige Abfassung der Arbeit 36
Curriculum Vitae 37
Danksagung 39
|