Confronted with the global demographic changes and the increasing pressure on modern healthcare system, there has been a surge of developing new technology platforms in the past decades. Droplet microfluidics is a prominent example of such technology platforms, which offers an efficient format for massively parallelized screening of a large number of samples and holds great promise to boost the throughput and reduce the costs of modern biomedical activities. Despite recent achievements, the realization of a compact and generic screening system which is suited for resource-limited settings and point-of-care applications remains elusive.
To address the above challenges, the dissertation focuses on the development of a compact multifunctional droplet micro-magnetofluidic system by exploring the advantages of magnetic in-flow detection principles. The methodologies behind a novel technique for biomedical applications, namely, magnetic in-flow cytometry have been put forth, which encompass magnetic indexing schemes, quantitative multiparametric analytics and magnetically-activated sorting. A magnetic indexing scheme is introduced and intrinsic to the magnetofluidic system. Two parameters characteristic of the magnetic signal when detecting magnetically functionalized objects, i.e. signal amplitude and peak width, providing information which is necessary to perform quantitative analysis in the spirit of optical cytometry has been proposed and realized. Magnetically-activated sorting is demonstrated to actively select individual droplets or to purify a population of droplets of interest. Together with the magnetic indexing scheme and multiparametric analytic technique, this functionality synergistically enables controlled synthesis, quality administration and screening of encoded magnetic microcarriers, which is crucial for the practical realization of magnetic suspension arrays technologies. Furthermore, to satisfy the needs of cost-efficient fabrication and high-volume delivery, an approach to fabricate magnetofluidic devices on flexible foils is demonstrated. The resultant device retains high performance of its rigid counterpart and exhibits excellent mechanical properties, which promises long-term stability in practical applications.
Identifer | oai:union.ndltd.org:DRESDEN/oai:qucosa:de:qucosa:20535 |
Date | 16 August 2016 |
Creators | Lin, Gungun |
Contributors | Schmidt, Oliver G., Cuniberti, Gianaurelio, Technische Universität Chemnitz |
Source Sets | Hochschulschriftenserver (HSSS) der SLUB Dresden |
Language | English |
Detected Language | English |
Type | doc-type:doctoralThesis, info:eu-repo/semantics/doctoralThesis, doc-type:Text |
Rights | info:eu-repo/semantics/openAccess |
Page generated in 0.0027 seconds