1 |
[en] NEW ANALYTICAL APPROACHES FOR SPECTROFLUORIMETRIC DETERMINATION OF CYCLOFENIL AFTER PHOTOCHEMISTRY DERIVATION AND FOR STROBILURIN VOLTAMMETRIC DETERMINATION USING MERCURY ELECTRODE AND CU (II) VOLTAMMETRIC DETERMINATION USING BISMUTH FILM ELECTRODE / [pt] NOVAS ABORDAGENS ANALÍTICAS PARA A DETERMINAÇÃO DE CICLOFENIL POR FLUORIMETRIA APÓS DERIVAÇÃO FOTOQUÍMICA E PARA AS DETERMINAÇÕES VOLTAMÉTRICAS DE ESTROBILURINAS UTILIZANDO O ELETRODO DE MERCÚRIO E DE CD (II), CU (II) E PB (II) UTILIZANDO O ELETRODO DE FILME DE BISMUTOWAGNER FELIPPE PACHECO 17 February 2009 (has links)
[pt] Neste trabalho, novas abordagens analíticas para a determinação
seletiva de ciclofenil, estrobilurinas (azoxistrobina e dimoxistrobina) e Cu (II)
foram desenvolvidas. A determinação do ciclofenil foi feita por meio do sinal
fluorescente obtido após sua fotoderivação com UV. Para se alcançar máxima
fluorescência (250/410 nm), as condições ótimas para derivação fotoquímica e
para medição espectrofluorimétrica foram otimizadas de forma univariada e, em
seguida, por meio da otimização multivariada (planejamento composto central).
Os valores otimizados mais críticos foram 2 h de tempo de exposição ao UV em
meio metanol/tampão Britton-Robinson pH 10,5. Um mecanismo para a reação
fotoquímica foi proposto, com a indicação da estrutura de dois prováveis
fotoderivados. Nestas condições, a resposta analítica teve comportamento linear e
homoscedástico (r2 = 0,9951), sendo o limite de detecção (LQ) e o limite de
quantificação (LQ) para determinação espectrofluorimétrica respectivamente
iguais a 1,1 x 10-8 mol L-1 e 3,7 x 10-7 mol L-1. A faixa linear do método se
estendeu até 8 x 10-5 mol L-1. Esta metodologia se mostrou adequada para a
determinação do ciclofenil em formulações farmacêuticas com recuperações de
98,3 ± 3,9 %. Para as determinações em urina, se fez necessário o uso da
separação cromatográfica em fase líquida. Nesse caso, o sinal fluorescente de um
dos picos do derivado fotoquímico (tempo de retenção em 7.0 min) foi escolhido.
Eficiente separação do sinal da matriz foi obtida em coluna C-18 usando modo de
eluição gradiente e fase móvel contendo metanol/água. O uso de padrão interno
(acetato de m-toluil) permitiu recuperações de 99,5 ± 3,4 % do analito em urina
fortificada. No caso da determinação seletiva de Cu (II) por voltametria de
PUC-Rio - Certificação Digital Nº 0410388/CA
redissolução anódica utilizando eletrodo de filme de bismuto, foi necessário
inicialmente resolver a interferência entre os picos de redissolução do Cu (II) e do
Bi (III), já que ambos aparecem em aproximadamente +80 mV. A interferência foi
contornada com a adição de peróxido de hidrogênio, que provocou o
deslocamento do potencial de redissolução do Cu (II) para +200 mV,
provavelmente devido à deposição de espécies oxidadas de cobre no filme de
bismuto. As condições experimentais e instrumentais foram otimizadas, de forma
univariada, para se obter um máximo de sinal analítico do Cu (II) sendo, para tal,
escolhidas as seguintes condições: potencial de deposição em -1400 mV,
velocidade de varredura de 400 mV s-1, 100 mV de amplitude, 120 s de tempo de
deposição e modo de varredura de onda quadrada. Nestas condições, foi
encontrado uma resposta analítica linear (R2 = 0,9927) e homoscedástica entre a
concentração de Cu (II) e a altura do pico de redissolução. O método foi aplicado
em amostras de cachaça onde se obteve valores de LD e de LQ respectivamente
iguais a 2,9 x 10-7 mol L-1 e 9,5 x 10-7 mol L-1 respectivamente. Os resultados
obtidos destas determinações foram comparados com os provenientes de outra
técnica (espectrofotometria de absorção atômica com forno de grafite) e não foi
encontrada nenhuma diferença estatística entre estas duas técnicas. A viabilidade
da determinação simultânea de Cd (II) e de Pb (II) também foi estudada e uma
condição experimental de compromisso estabelecida. Para a estrobilurinas, um
método voltamétrico sensível e seletivo foi desenvolvido para a aplicação em
alimentos, água de rio e suco industriais. As condições instrumentais foram
desenvolvidas de modo a se obter um máximo de sinal analítico para
determinação simultânea de azoxistrobina e dimoxistrobina no eletrodo de
mercúrio. As condições otimizadas foram: potencial inicial em -300 mV, 150 mV
de amplitude, velocidade de varredura de 300 mV s-1, modo de varredura de onda
quadrada. Nestas condições foram observados sinais para a azoxistrobina e para a
dimoxistrobin / [en] In this work, analytical approaches for the selective determinations of
cyclofenil, stroblilurins (azoxystrobin and dimoxystrobin) and Cu (II) have been
developed. The determination of cyclofenil was made by taking advantage of their
fluorescent photoderivative. In order to achieve maximum fluorescence (250/410),
the best conditions for derivatization and for the analytical measurements have
been optimized firstly by univariate approach and then by a multivariate way
(central composite design). The optima conditions were: 2 h of UV exposition
time and solvent system composed by methanol/Britton-Robinson buffer pH 10.5.
A mechanism for this photochemical reaction has been proposed, with the
indication of the structure of two most probable photoderivades. In such
conditions, the analytical response presented a linear and homocedastic behavior
(r2=0,9951) up to 8 x 10-5 mol L-1. Limit of detection (LQ) and limite of
quantification (LQ) for the spectrofluorimetric determination were respectively
1.1 x 10-8 mol L-1 and 3.,7 x 10-7 mol L-1. The methodology was proven to be
adequate to quantify cyclofenil in pharmaceutical formulations with mean
recovery of 98.3 + 3.9%. For the determination of cyclofenil in urine, it was
necessary the use of a separation technique (HPLC). In such case, the fluorescent
signal from one of its photochemical derivatives (7.0 min retention time) was
chosen. Efficient separation from the matrix background was achieved using a C-
18 column, gradient elution using methanol/water system. The use of an internal
standard (m-toluil acetate) allowed recoveries of 99.5 + 3.4 % for cyclofenil
fortified urine samples. In the case of the selective anodic stripping voltammetric
determination of Cu (II) using bismuth film electrode, the overlapping between
Cu (II) and Bi (III) needed to be eliminated, since both species present redissolution
peaks near +80 mV. The interference was solved by the addition of
PUC-Rio - Certificação Digital Nº 0410388/CA
hydrogen peroxide in the working cell which caused the shifting of Cu (II) redissolution
peak to +200 mV, probably caused by the deposition of oxidated
copper species on the electrode surface. The experimental and instrumental
conditions have been optimized in an univariated way in order to achieve a
maximum analytical signal for Cu (II). As a result, the optimized conditions were:
-1400 mV of deposition potential, 400 mV s-1 scan rate, 100 mV amplitude. In
such conditions, a linear and homoscedastic behavior (r2 = 0,9927) analytical
response was found. The method has been applied for sugarcane spirit samples
where values of 2.9 x 10-7 mol L-1 and 9.5 x 10-7 mol L-1 were achieved
respectively for LD and LQ. The results from such determinations had been
compared with the ones from another technique (graphite furnace atomic
absorption spectrometry) with statistically similar results. The simultaneous
determination of Cd (II) and Pb (II) was also achieved using a compromised
condition. For the strobilurins, a sensible and selective voltammetric method was
developed to be applied in foods, river water and industrial juices. The
experimental conditions have been developed in order to achieve a maximum
analytical signal for the simultaneous determination of azoxistrobin and
dimoxistrobin in the hanging mercury drop electrode. The optimized conditions
were: deposition potential at -300 mV, 150 mV amplitude, 300 mVs-1 using the
square wave mode. In such conditions, the azoxystrobin and the dimoxystrobin
peaks were found at – 928 mV and -650 mV respectively. Linear analytical
response (r2>0,99) and homoscedastic behavior have been observed. LD and LQ
where respectively 3.6 x 10-7 mol L-1 and 1,2 x 10-6 mol L-1 for azoxystrobin and
7.1 x 10-7 mol l-1 and 2.4 x 10 -6 mol L-1 for dimoxystrobin. Enough selectivity
was achieved in complex samples and no interferences from others strobilurins
(piraxystrobin and picoxystrobin) was found. A simple mechanism for the redox
process at the mercury electrode was proposed.
|
2 |
[en] DEVELOPMENT OF ELECTROANALYTICAL METHODS FOR THE DETERMINATION OF SIBUTRAMINE IN PHARMACEUTICAL FORMULATIONS AND IN PROTEIN-RICH FOOD COMPLEMENTS / [pt] DESENVOLVIMENTO DE METODOLOGIAS ELETROANALÍTICAS PARA A DETERMINAÇÃO DE SIBUTRAMINA EM FÁRMACOS E EM ALIMENTOS PROTÉICOSJULIANA MACHADO DE CARVALHO 22 October 2009 (has links)
[pt] O cloridrato de sibutramina está entre os sacietógenos mais prescritos para
o controle da obesidade e tem sido adicionado, de forma não declarada, a
alimentos protéicos para esportistas. Nenhuma metodologia analítica é descrita
em compêndios oficiais para a determinação de sibutramina. O presente
trabalho objetiva o desenvolvimento e a validação de métodos analíticos
baseados na voltametria de pulso diferencial (DPV) e na voltametria de onda
quadrada (SWV) para quantificação do cloridrato de sibutramina em fármacos e
em alimentos protéicos. A influência dos parâmetros eletroquímicos foi
estudada para selecionar as melhores condições de trabalho como: eletrólito
suporte (tampão Mcllvaine) pH (4,0), velocidade de varredura (40 mV s(-1) para
DPV e 240 mV s(-1) para SWV) e amplitude de pulso (50 mV). Não foi possível
pré-concentrar a sibutramina no eletrodo de trabalho (gota de mercúrio). O
comportamento eletroquímico desta substância no eletrodo de mercúrio também
foi investigado a partir de estudos com voltametria cíclica. A redução da
sibutramina é aparentemente reversível envolvendo a participação de 1 próton e
1 elétron. O potencial de pico da sibutramina é cerca de - 100 mV (Ag/AgCl 3
mol L-1). No processo de validação foram avaliadas as faixas de resposta linear
(do limite de quantificação até 33,3 mg L(-1) em ambos os métodos), o limite de
detecção (0,4 para DPV e 0,3 mg L(-1) para SWV), e o limite de quantificação
(1,4 para DPV e 1,1 mg L(-1) para SWV). A precisão do método foi satisfatória.
Testes de recuperação indicaram 90,4 % em ambos os métodos. O desempenho
das técnicas voltamétricas foi comparado estatisticamente com a técnica CLAE
e não houve diferença significativa entre os resultados obtidos. / [en] Sibutramine hydrochloride is among the most prescribed satiety inducer
for obesity control and it has also been added, in a fraudulent way, in proteinrich
food complement for athletes. There is no analytical method for
sibutramine described in the official literature. The goal of this work is to
develop and validate electroanalytical methods for the determination of
sibutramine in pharmaceuticals and in protein-rich food complements.
Differential pulse voltammetry (DPV) and square wave voltammetry (SWV)
were the techniques employed. The influence of electrochemical parameters
was studied in order to select the best working conditions such as supporting
electrolyte (Mcllvaine buffer), pH (4.0), scan rate (40 mV s(-1) for DPV and 240
mV s(-1) for SWV) and pulse amplitude (50 mV). The pre-concentration of
sibutramine in the working electrode (mercury drop) was not possible. The
electrochemical behavior in the mercury electrode was also investigated using
cyclic voltammetry. The reduction of sibutramine is probably reversible and
involves of 1 proton and 1 electron. The peak potential of sibutramine is about -
100 mV (Ag/AgCl3 mol L(-1)). In the validation process, the linear range of the
analytical response (from the limit of quantification to 33.3 mg L-1, in both
methods), the limit of detection (0.6 for DPV and 0.4 mg L(-1) for SWV) and the
limit of quantification (1.8 for DPV and 1.1 mg L(-1) for SWV) were evaluated.
Satisfactory precision was achieved. Recovery tests produced results of 90.4%
for both methods. These results were compared with the ones achieved by
HPLC and no significant statistical difference was found between them.
|
3 |
[en] DEVELOPMENT AND COMPARISON OF VOLTAMMETRIC METHODS FOR THE DETERMINATION OF CYCLOFENIL AND PRIMAQUINE IN PHARMACEUTICAL FORMULATIONS AND IN URINE / [pt] DESENVOLVIMENTO E COMPARAÇÃO DE MÉTODOS VOLTAMÉTRICOS PARA A DETERMINAÇÃO DE CICLOFENIL E PRIMAQUINA EM MEDICAMENTOS E EM URINAWAGNER FELIPPE PACHECO 13 July 2004 (has links)
[pt] Neste trabalho foram desenvolvidas metodologias
eletroanalíticas baseadas
na voltametria adsortiva de redissolução catódica com
varredura de potencial de
onda quadrada e de pulso diferencial para a determinação do
ciclofenil em urina e
em formulações farmacêuticas, e na voltametria anódica com
varredura de
potencial de onda quadrada e de pulso diferencial para a
determinação da
primaquina em formulações farmacêuticas. Comparando-se as
duas técnicas de
aquisição, a voltametria de onda quadrada foi escolhida
para realizar as
determinações do ciclofenil em urina e em formulação
farmacêutica por se
mostrar, além de uma técnica mais rápida, maior
sensíbilidade. Para a
determinação da primaquina a melhor técnica foi à
voltametria de pulso
diferencial.
As condições experimentais que possibilitaram um melhor
desempenho
analítico em termos da obtenção do menor limite de detecção
e maior
reprodutibilidade das leituras foram otimizados. No caso do
ciclofenil, o
composto mostrou ser instável no meio aquoso e orgânico
para determinação
voltamétrica, com sistemática diminuição da corrente
faradaica. Assim, fez-se uso
das propriedades eletroquímicas de um derivado estável
obtido pela derivatização
fotoquímica do composto (o composto foi irradiado com
irradiação UV por 45
minutos em um reator fotoquímico), a corrente máxima obtida
apresentou um
potencial a -1,28 V. As condições experimentais que
possibilitaram este sinal
foram obtidas com 30 s de tempo de deposição do analito
sobre o eletrodo de
mercúrio, solução Britton-Robbinson pH 9,0 como eletrólito
de suporte, potencial
de acumulação de -0,9 V, amplitude de 250 mV, incremento de
potencial de 2,0
mV no modo de onda quadrada.
Desta forma foi obtido limite de detecção da ordem de 10-8
mol L-1, faixa
linear dinâmica de 2 ordens de grandeza, condições estas
que possibilitaram a
determinação do composto tanto em formulações farmacêuticas
(Menopax)
como em amostra de urina enriquecida com o analito de
interesse. Foram feitos
testes com subtâncias concomitantes da formulação
farmacêutica e constatou-se
que não existia problema com interferência na análise nesse
tipo de amostra. Nas
determinações em urina não houve a necessidade de se fazer
tratamento prévio da
amostra, apenas a irradiação UV para estabilizar o
composto, as análises foram
realizadas com o método de adição de analito, de forma a se
corrigir a
interferência de matriz. Em ambos os casos os resultados
obtidos se encontram na
faixa de 93,6 a 106,5 por cento, dentro da faixa de recuperação
aceitável para este tipo de
problema analítico segundo estabelecido pela Farmacopéia
dos Estados Unidos da
América.
Usou-se ainda os resultados provenientes das voltametrias
de pulso
diferencial, de voltametria de onda quadrada e da
voltametria cíclica para se obter
informações sobre o processo redox que ocorria no eletrodo
de trabalho.
Constatou-se que a reação é reversível, com um processo de
controle adsortivo,
sendo que apenas o reagente adsorve no eletrodo de
trabalho, não ocorrendo
adsorção do produto da reação de redução. Constatou-se
também que o processo
envolve a transferência de apenas 1 elétron, e que a reação
não possuía
contribuição cinética.
No caso da primaquina o sinal de redução ocorre em região
anódica (0,592
V), foi portanto necessário utilizar um eletrodo de carbono
para se poder fazer a
determinação da primaquina. Visando a possibilidade do uso
do eletrodo de
mercúrio, tentou-se fazer uso da formação de derivados
complexos da primaquina
com vanádio, com cobre ou com íon iodeto (complexo de
tranferência de carga),
porém não foram observados sinal da reação redox na janela
de potencial
procurada. Nas condições experimentais otimizadas não se / [en] In this work, square-wave and differential pulse
voltammetric methods were
developed for the determination of cyclofenil and
primaquine in pharmaceutical
formulations and in urine samples. The use of the square-
wave acquisition
technique was found to enable better sensitivity and faster
analysis time compared
to the differential pulse technique.
Experimental and instrumental conditions were optimized to
allow the best
analytical performance in terms of limit of detection and
repeatability of the
readings. In the case of cyclofenil, its unstable behavior
in aqueous and organic
solvents, with systematic decreasing of analyte current
signal, makes impossible
any voltammetric determination. As an alternative way, the
electrochemical
properties of a stable photochemical derivative of
cyclofenil was used (the
compound was irradiated with UV radiation for 45 min) with
maximum current at
-1,28 V. This analyte photoderivative could also be
accumulated in the working
electrode. The experimental conditions that allowed the
maximum current was a
30 s of deposition time at the mercury electrode, Britton-
Robbinson (pH 9,0)
supporting electrolyte, accumulation potential of -0,9 V,
amplitude of pulse of
250 mV, scan increment 2,0 mV.
These optimized conditions allowed a limit of detection of
10-8 mol L-1 and
dynamic linear range of 2 orders of magnitude to be
achieved. These analytical
figures of merit made possible the determination of
cyclofenil either in a
pharmaceutical formulation (Menopax) and in urine samples
spiked with the
analyte of interest. The potential interferences from
concomitant substances used
in the pharmaceutical formulation were also evaluated. For
the analyte
determination in urine, only UV irradiation of sample was
necessary, in order to
obtain stable cyclofenil derivative. The analyte addition
method was used to
analyze urine in order to minimize matrix interferences.
Recovery results for the
analysis of Menopax and for the analysis of urine were
between 96,5 and 107,6
percent, within the acceptable recovery range established by the
United States
Pharmacopoeia.
Information concerning the analyte redox reaction and
electrode processes
was also obtained from differential pulse voltammetry,
square-wave voltammetry
and cyclic voltammetry. It was verified that the cyclofenil
photoderivative
eletrochemical reaction is reversible with adsorption of
only the reagent on the
surface of the electrode. The adsorption of the
electrochemical reduction product
does not occur. It was also verified that the process
involves the transference of
only one electron, and there is no kinetics contribution in
the reaction.
In the case of the primaquine the analyte reduction occurs
in anodic region
(0,592 V), therefore, it was necessary the use the carbon
glass electrode to allow
the determination of this analyte. The pre-concentration of
the analyte in the
working carbon glass electrode was also not attained with
the experimental
conditions used.
Several attempts were made to make possible the use of the
mercury
electrode, including the formation of charge transfer
complexes with iodine and
complexation with vanadium. However no success was obtained.
Using the carbon glassy electrode and DPV technique the
determination of
primaquine in pharmaceutical formulations and urine was
performed with average
recovery of 101.7 percent. Limit of detention of 1,0 x 10-6 mol L-
1 was obtained.
|
4 |
[en] DEVELOPMENT OF ELECTROANALYTICALS METODOLOGY FOR DETERMINATION OF ZIDOVUDINE, LAMIVUDINE AND ZALCITABINE / [pt] DESENVOLVIMENTO DE METODOLOGIAS ELETROANALÍTICAS PARA A DETERMINAÇÃO DE ZIDOVUDINA, LAMIVUDINA E ZALCITABINAKATIA CHRISTINA LEANDRO ANTUNES 14 October 2004 (has links)
[pt] O comportamento eletroquímico e o desenvolvimento de metodologias eletroanalíticas para a determinação de zidovudina (AZT), lamivudina (3TC) e zalcitabina (DDC) em medicamentos foram investigados no presente trabalhoatravés de técnicas voltamétricas. Estas substâncias possuem atividade antiretroviral, sendo utilizadas no tratamento de indivíduos portadores de HIV. AZT, 3TC e DDC
sofrem redução na superfície do eletrodo de mercúrio. Ainfluência dos parâmetros eletroquímicos, tais como, velocidade de varredura, amplitude, natureza do eletrólito suporte e pH foram estudados para selecionar as melhores condições para quantificar estes princípios ativos em formulações farmacêuticas. A redução da zidovudina ocorre em apenas uma etapa, sendo esta atribuída à redução do grupo azida (-N3), originando em um pico de potencial de-970 mV vs Ag/AgCl (3 mol L-1), utilizando tampão fosfato,
pH 8,0. A faixa linear encontrada foi de 0,25 a 1,25 mg L-1. O mecanismo de redução da zidovudina envolve a participação de 4 elétrons e 2 prótons. Adicionalmente novos estudos eletroquímicos foram desenvolvidos para a determinação de lamivudina e de zalcitabina através da voltametria em pulso diferencial em tampão Clark-Lubs, pH 2,0, onde se obteve um pico no potencial de -1160 e -1180 mV, respectivamente. Lamivudina obteve faixa linear de 2,28 a 8,85 mg L-1 e zalcitabina de 10,0 a 23,3 mg L-1. O mecanismo de redução destas drogas envolve a participação de números iguais de prótons e elétrons. Todas as metodologias desenvolvidas também foram validadas. / [en] The electrochemical behaviour and the development of an analytical methodology for the zidovudine (AZT), lamivudine (3TC) and zalcitabine (DDC)determination in drugs using voltammetrics techniques were the objectives of the current study. These drugs have been extensively used in the treatment of HIV patients because its antiretroviral activity. The AZT, 3TC and DDC are reduced at a hanging mercury drop electrode (HMDE). The influence of electroanalytical parameters such as, scan rate, amplitude, nature of the support electrolyte and pH in the signal was verified to select the best conditions for quantifying these drugs in pharmaceutical forms. The obtained results had showed AZT, in solution of pH 8.0 of phosphate buffer, with only one reduction in -970 mV vs Ag/AgCl (3 mol L-1). This peak is due to reduction of the azido group present on zidovudine molecule. The linear range of standards is from 0.25 to 1.25 mg L-1. The electrodic mechanism involving the participation of 4 electrons and 2 protons in the reduction of AZT molecules. News studies were development for lamivudine and zalcitabine determination by differential pulse voltammetry in solution of pH 2.0 of Clark-Lubs buffer, with only one reduction in - 1160 and -1180 mV, respectively. Lamivudine has been the linear range of standards from 2.28 to 8.85 mg L-1 and
zalcitabine from 10.0 to 23.3 mg L-1. The mechanism of these drugs involving the participation of the same number of electrons and protons. All these analytical methods were validated.
|
5 |
[pt] DESENVOLVIMENTO DE MÉTODOS ELETRO-ANALÍTICOS USANDO SENSORES MODIFICADOS COM NANOMATERIAIS PARA DETERMINAÇÃO DE PRIMAQUINA, INHHQ, TIOMERSAL E CREATININA / [en] DEVELOPMENT OF ELECTRO-ANALYTICAL METHODS USING SENSORS MODIFIED WITH NANOMATERIALS FOR DETERMINATION OF PRIMAQUINE, INHHQ, THIOMERSAL AND CREATININEMARLIN JEANNETTE PEDROZO PEÑAFIEL 28 December 2020 (has links)
[pt] O objetivo deste trabalho foi o desenvolvimento de métodos eletro-analíticos utilizando eletrodos modificados com nanomateriais a base de carbono para determinação de analitos de interesse biológico (creatinina) e farmacológico (primaquina, INHHQ, tiomersal).
A determinação de primaquina (antimalárico) e de INHHQ (potencial medicamento para o tratamento de doenças neurodegenerativas) foi proposta usando eletrodo de carbono vítreo (GC) modificado com nanotubos de carbono de paredes múltiplas (MWCNTs) o qual promoveu uma melhora significativa na corrente faradaica e na resolução de pico obtida usando voltametria de onda quadrada (SWV) em tampão Britton-Robinson (BR) 0.02 mol L-1; pH 7.00 e KCl (0.25 mol L-1). Embora vários métodos para quantificar primaquina estejam reportados, muitos deles dependem de procedimentos considerados complexos e trabalhosos, portanto, a modificação proposta de GC usando MWCNTs é comparativamente fácil, levando a uma modificação robusta que produz resultados repetitivos, fornecendo também amplificação do sinal eletroquímico devido à área ativa aumentada e a taxa de transferência eletrônica aprimorada. Os GC/MWCNTs também forneceram, ao trabalhar em um pH adequado, maior resolução na detecção das diferentes etapas eletroquímicas envolvidas no processo redox da primaquina.
O eletrodo permitiu o limite de detecção (LOD) de 28 nmol L-1 para a primaquina e, quando associado a um procedimento simples baseado na extração líquido-líquido (LLE) e cromatografia em camada fina (TLC), foi alcançada a seletividade e a capacidade de determinar os níveis mais baixos esperados em amostras de urina. No caso do INHHQ, um método para quantificação por SWV foi relatado pela primeira vez. Além do uso do eletrodo GC/MWCNTs um prévio procedimento de TLC, permitiu a seletividade necessária para determinar o INHHQ no tecido cerebral de ratos Wistar. O LOD foi de 0.85 μmol L-1 e a faixa dinâmica linear cobriu duas ordens de magnitude (10-6 a 10-5 mol L-1).
Os pontos quânticos de grafeno (GQDs) foram escolhidos para modificar quimicamente os eletrodos utilizados para determinar tiomersal e creatinina. GQDs são fragmentos nanométricos de grafeno (presentes em dispersões como monocamada, poucas camadas ou multicamadas) onde o transporte de elétrons é confinado em suas dimensões. GQDs são estruturas baseadas em carbono que são biocompatíveis e apresentam baixa toxicidade. A pesquisa apresenta uma nova abordagem para a determinação de tiomersal (conservante em vacinas e cosméticos) em vacinas contra influenza usando GQDs como modificador do eletrodo de GC e explorando o efeito sinérgico entre GQDs, radiação visível e a diferença de potencial aplicado. Este sistema promove a oxidação do tiomersal, produzindo um aumento significativo na resposta (par redox Hg/Hg2positivo) em termos de intensidade de corrente usando SWV e tampão BR (0.02 mol L-1; pH 4.00 com KCl a 0.25 mol L-1). O LOD foi de 0.85 μmol L-1 e a faixa dinâmica linear cobriu três ordens de grandeza (10-6 a 10-4 mol L-1). Para a creatinina (um biomarcador para a disfunção renal), um composto de GQDs-cobre foi usado para modificar um eletrodo de ouro. A determinação da creatinina foi realizada utilizando SWV (tampão BR 0.02 mol L-1; pH 4.00 com KCl a 0.25 mol L-1). O método contou com a diminuição do pico de oxidação Cu/Cu2positivo na presença de creatinina, permitindo LOD de 50 nmol L-1 e faixa dinâmica linear, cobrindo três ordens de magnitude (10-6 a 10-4 mol L-1). A resposta SWV da creatinina foi atribuída à formação do complexo creatinina-Cu2+, sendo razoavelmente seletiva em relação a este analito. Todos os procedimentos desenvolvidos por voltametria foram adequadamente comparados com métodos de referência, sendo estatisticamente compatíveis. / [en] The goal of this work was the development of electro-analytical methods using electrodes modified with carbon-based nanomaterials for determination of analytes of biological (creatinine) and pharmacological interest (primaquine, INHHQ, thiomersal). The determination of primaquine (antimalarial) and INHHQ (potential drug for treatment of neurodegenerative diseases) was proposed using glassy carbon (GC) electrode modified with multi-walled carbon nanotubes (MWCNTs) that promoted significant improvement in faradaic current and better peak resolution obtained using square-wave voltammetry (SWV) in Britton-Robinson (BR) buffer 0.02 mol L-1; pH 7.00 and KCl (0.25 mol L-1). Although several methods to quantify primaquine are reported, many of these depend on procedure considered complex and laborious, therefore, the proposed GC modification using MWCNTs is comparatively easy, leading to a robust modification that produce repetitive results also providing amplification of the electrochemical signal due to the increased active area and improved electron-transfer rate. The GC/MWCNTs also provided, when working on a proper pH, higher resolution in detecting the different electrochemical steps involved in the redox process of primaquine. The electrode enabled limit of detection (LOD) of 28 nmol L-1 for primaquine and, when associated with a simple developed procedure, relying of liquid-liquid extraction (LLE) and thin-layer chromatography (TLC), it was achieved the selectivity and the capability to determine the lower levels expected in urine samples. In the case of INHHQ, a method for quantification by SWV was reported for the first time. In addition to the use of GC/MWCNTs electrode, a previous TLC procedure, enabled the selectivity required to determine INHHQ in brain tissue from Wistar rats. The LOD was 0.85 mol L-1 and the linear dynamic range covered two orders of magnitude (10-6 to 10-5 mol L-1).
Graphene quantum dots (GQDs) was chosen to chemically modify electrodes used to determine thiomersal and creatinine. GQDs are nanometric fragments of graphene (present in dispersions as monolayer, few layer or multilayer) where electron transport is confined in their dimensions. GQDs are carbon-based structure that are bio-compatible and present low toxicity. The research presents a new approach for the determination of thimerosal (preservative in vaccines and cosmetics) in influenza vaccines using GQDs as a modifier of the GC electrode and exploring the synergistic effect between GQDs, visible radiation and the applied potential difference. This system promotes thiomersal oxidation, producing a significant increase in response (Hg/Hg2 positive redox pair) in terms of current intensity using SWV and BR buffer (0.02 mol L-1; pH 4.00 with KCl at 0.25 mol L-1). The LOD was 0.85 μmol L-1 with linear dynamic range covering three orders of magnitude (10-6 to 10-4 mol L-1). For creatinine, a biomarker for the renal dysfunction, a GQDs-copper composite was used to modify a gold electrode. Determination of creatinine was made using SWV (BR buffer 0.02 mol L-1; pH 4.00 with KCl at 0.25 mol L-1). The method relied on the decreasing of the Cu2+/Cu oxidation peak in the presence of creatinine, enabling LOD of 50 nmol L-1 and linear dynamic range covering three orders of magnitude (10-6 to 10-4 mol L-1). The SWV response of creatinine was attributed to the formation of creatinine-Cu2 positive complex being reasonably selective towards this analyte. All procedures developed by voltammetry were adequately compared with reference methods, being statistically compatible.
|
6 |
[en] LOW-TEMPERATURE SINTERING OF TITANIA USED IN PHOTOCATALYTIC REACTIONS / [pt] SINTERIZAÇÃO A BAIXAS TEMPERATURAS DA TITÂNIA USADA EM REAÇÕES FOTOCATALÍTICASANNA LUISA WERNECK RUOTOLO MIGUEL 16 November 2021 (has links)
[pt] A busca pela diminuição da dependência de combustíveis fósseis faz com
que a ciência avance, diariamente, na utilização de combustíveis ecológicos,
como o H2. Uma das formas de sua obtenção é através da fotocatálise. Esse
processo consiste em uma reação catalítica com o uso de energia, na forma de
luz. A fotólise da água é amplamente utilizada, principalmente com luz solar como
fonte luminosa, que é abundante e reduz os custos de sua produção. O
desempenho desta reação depende da posição das bandas de condução (BC) e
valência (BV) do fotocatalisador. O TiO2 é utilizado como fotocatalisador em
diversas reações, inclusive para a produção de H2. Os fotocatalisadores utilizados
na forma de pós nanométricos apresentam dificuldade de separação após a
reação. A transformação do pó em um material compacto é uma alternativa para
retirá-lo do meio reacional evitando perdas e custos com separação. Desse modo,
a compactação do pó é uma alternativa para facilitar sua reciclagem. O principal
método de sua produção é pelo processo de sinterização, que envolve
temperaturas elevadas (geralmente, 75 por cento do ponto de fusão do material) e longo
tempo, podendo durar até dias. Para diminuir os gastos energéticos, o processo
de sinterização a frio é uma opção, que consiste na densificação do material com
uso de pressão e um solvente (aquoso ou não) e, temperaturas de sinterização
de até 500 Graus C. O objetivo do estudo consistiu na produção de pastilhas de TiO2,
comercial, e P25, através de uma variação do método de sinterização a frio, onde
aplicou-se a pressão no pó, junto com o solvente, antes de seu tratamento térmico.
As pastilhas produzidas foram caracterizadas pelas técnicas de TGA/DSC, XRD,
MEV, CV, e DRS. / [en] The quest to reduce dependence on fossil fuels makes science advance,
daily, in the use of ecological fuels, such as H2. One of the ways to obtain it is
through photocatalysis. This process consists on a catalytic reaction using energy,
in the form of light. Water photolysis is widely used, mainly with sunlight as a light
source, which is abundant and reduces production costs. The performance of this
reaction depends on the position of the conduction (CB) and valence (VB) bands
of the photocatalyst. TiO2 is used as a photocatalyst in several reactions, including
the production of H2. Photocatalysts used in the form of nanometric powders have
difficulty in separating after the reaction. The transformation of the powder into a
compact material is an alternative to remove it from the reaction medium, avoiding
losses and costs with separation. Thus, the compaction of the powder is an
alternative to facilitate its recycling. The main method of its production is through
the sintering process, which involves high temperatures (generally 75 percent of the
material s melting point) and a long time, which can last up to days. To reduce
energy costs, the cold sintering process is an option, which consists of densifying
the material using pressure and a solvent (aqueous or not) and sintering
temperatures of up to 500 C degrees. The aim of the study was the production of
commercial TiO2 and P25 pellets, through a variant of the cold sintering method,
where pressure was applied to the powder, with the proper solvent, before the heat
treatment. The pellets produced were characterized by the techniques of
TGA/DSC, XRD, SEM, CV, and DRS.
|
7 |
[en] DEVELOPMENT AND COMPARATIVE STUDY OF SPECTROFLUORIMETRIC AND VOLTAMMETRIC METHODOLOGIES FOR THE DETERMINATION OF THALIDOMIDE IN ONE COMMERCIAL FORMULATION, URINE AND BLOOD SERUM / [pt] DESENVOLVIMENTO E ESTUDO COMPARATIVO DE METODOLOGIAS ESPECTROFLUORIMÉTRICA E VOLTAMÉTRICA PARA A DETERMINAÇÃO DE TALIDOMIDA EM UM FÁRMACO, URINA E SORO SANGÜÍNEOCARLOS EDUARDO CARDOSO 21 July 2003 (has links)
[pt] No presente trabalho foram desenvolvidas duas metodologias
analíticas para a determinação de talidomida, um composto
de reconhecida importância farmacológica. As metodologias
espectrofluorimétrica e voltamétrica desenvolvidas foram
comparadas em termos de desempenho analítico.As
características fluorescentes e eletroquímicas da
talidomida foram estudadas para que se encontrassem
condições experimentais que fornecessem máximo sinal
fluorescente do analito em solução na temperatura ambiente e
possibilidade de pré-concentração do analito no eletrodo de
mercúrio.Nas condições experimentais otimizadas para a
talidomida, limites de detecção compreendidos entre 10-6 e
10-9 g L-1 foram obtidos para o método
espectrofluorimétrico e voltamétrico, respectivamente.
Faixas lineares dinâmicas entre 2 e 4 ordens de grandeza
foram alcançadas dependendo do método e do interferente
presente na amostra. Esses parâmetros de mérito se mostraram
adequados para o problema proposto. O possível efeito
interferente de substâncias geralmente usadas em
associação com o analito foi estudado, e estratégias para
minimização das interferências foram desenvolvidas.
Enquanto a tetraciclina não interferiu no método
espectrofluorimétrico, o uso combinado de meio ácido e
irradiação UV da amostra foi necessária para permitir a
quantificação do analito em presença de sulfanilamida. Na
voltametria, as interferências da tetraciclina e da
sulfanilamida puderam ser compensadas pelo uso da
quantificação pelo método de adição do analito.
Nas determinações dos fluídos biológicos, o uso da extração
em coluna de sílica C18 mostrou-se bastante eficiente na
separação do analito dos interferentes da matriz para o
método espectrofluorimétrico e para o voltamétrico, no caso
do soro sangüíneo. Para a urina, apenas a clarificação da
amostra com sulfato de amônio foi suficiente no caso da
determinação voltamétrica.Os métodos desenvolvidos foram
testados na dosagem de talidomida presente em uma
formulação comercial e em amostras de urina e soro sangüíneo
enriquecidas com o analito. Para tal, utilizaram-se os
procedimentos de curva de calibração (espectrofluorimetria)
e método da adição do analito (voltametria). Em todos os
casos, as recuperações obtidas estiveram compreendidas na
faixa de 96,5 a 107,6 %, dentro da faixa de recuperação
estabelecida pela Farmacopéia dos Estados Unidos da América. / [en] In the present work two analytical methodologies were
developed aiming the determination of thalidomide, an
important pharmacological compound. The developed
spectrofluorimetric and the voltammetric based analytical
methodologies were compared in terms of analytical
performance. The thalidomide fluorescent and the
electrochemical characteristics were studied in order to
find experimental conditions for maximum fluorescence in
solution and at room temperature and to allow analyte pre-
concentration on the mercury electrode. Using the optimized
experimental conditions, limits of detection between
10-6 to 10-9 g L-1 were achieved respectively for the
spectrofluorimetric method and for the voltammetric method.
Dynamic linear ranges between 2 and 4 orders of magnitude
were obtained depending on the method utilized and the
interferent substances present in the sample. Those
parameters of merit were suitable for this proposed
analytical problem. The potential interference effect from
substances usually used in association with thalidomide,
were studied and strategies for the minimization of such
interferences were developed. While no interference in the
spectrofluorimetric method was observed for tetracycline,
the combined use of acidic medium and UV irradiation of the
samples was necessary to allow the analyte determination in
the presence of sulfanilamide. For the voltammetric method,
interferences from tetracycline and sulfanilamide could be
compensated by quantifying thalidomide using the analyte
addition method. For the determination in biological
fluids, the use a solid-liquid extraction on a C18 column
was found to be very effective for the analyte separation
and elimination of matrix interferences for the
spectrofluorimetric method and for the voltammetric method
developed for blood serum. For urine samples, a clean-up
step using ammonium sulfate was found to be sufficient for
the voltammetric determination of thalidomide.
The developed methodologies were tested by determining the
thalidomide content in a commercial pharmaceutical
formulation and in analyte spiked biological fluids using
calibration curves (spectrofluorimetric) and analyte
addition method (voltammetric). In all cases, the
recoveries were between the 96,5 and 107,6 %, within the
recovery range considered adequate according to the
United States Pharmacopoeia.
|
8 |
[en] DEVELOPMENT OF ELECTRO-ANALYTICAL METHODS USING ELECTROCHEMICAL CARBON-BASED SENSORS FOR DETERMINATION OF TRIFLOXYSTROBIN, GENTAMICIN AND LAPACHOL / [pt] DESENVOLVIMENTO DE MÉTODOS ELETROANALÍTICOS UTILIZANDO SENSORES ELETROQUÍMICOS BASEADOS NO CARBONO PARA DETERMINAÇÃO DE TRIFLOXISTROBINA, GENTAMICINA E LAPACHOLJOSEANY DE MORAES SANTOS ALMEIDA 20 December 2018 (has links)
[pt] O presente trabalho teve como objetivo o desenvolvimento de métodos eletroanalíticos empregando elétrodos de grafite-epóxi e diamante dopado com boro para a determinação do antibiótico sulfato de gentamicina, substâncias de interesse biológico (naftoquinonas) e do pesticida trifloxistrobina. O diamante dopado com boro (DDB) apresenta características atraentes de detecção eletroquímica que são úteis em aplicações analíticas baseadas em voltametria e amperometria. Possui uma ampla janela de potencial em soluções aquosas que permitiu a quantificação do fungicida trifloxistrobina (em +1744 mV versus Ag/AgCl(KClsat)) por voltametria de onda quadrada (SWV), em um tampão Britton-Robinson (0,04 mol L-1, pH 4,00 )/acetonitrila 70/30 por cento v/v. A curva de adição de analito foi obtida usando amplitude de pulso 40 mV, freqüência de 30 Hz, passo de potencial de 20 mV. O limite instrumental de detecção foi de 1,4 × 10-7 mol L-1 e a faixa linear dinâmica abrangeram três ordens de grandeza (10-7 a 10-5 mol L-1). As amostras foram analisadas com recuperações de cerca de 80 por cento em amostras de suco de laranja e de 92,4 a 104,0 por cento em amostras de água. Um estudo para avaliar potenciais interferentes foi realizado na presença de outros fungicidas. Estudos diagnósticos indicaram que a oxidação da trifloxistrobina em meio aquoso na superfície do DDB é irreversível, envolvendo duas etapas, cada um com dois elétrons. A degradação UV da trifloxistrobina foi avaliada utilizando o método eletroquímico proposto e a cinética de degradação estabelecida com meia-vida de 1.07 min. A amperometria por injeção em batelada (BIA do inglês batch-injection analysis) com o DDB (operando a +2000 mV) foi utilizada para desenvolver um método para a determinação de sulfato de gentamicina. A extração em fase sólida (SPE do inglês solid-phase extraction), utilizando um cartucho empacotado com um polímero impresso molecularmente, tendo a canamicina como molécula molde, foi utilizada para melhorar a seletividade. O tampão acetato (0,01 mol L-1; pH 4,4) foi utilizado como eletrólito suporte. A oxidação do sulfato de gentamicina ocorreu logo após a injeção do analito (60 microlitros) que entra em contato com a superfície do eletrodo. O cartucho de SPE foi muito eficiente na retenção do analito quando comparado com o polímero não-impresso correspondente (NIP). A curva analítica apresentou resposta linear, a precisão instrumental foi inferior a 3 por cento e o limite de quantificação foi de 2,7 × 10-6 mol L-1. A frequência analítica foi de 90 medições h-1 e nenhum efeito de memória foram observados entre as medidas sequênciais. O desempenho do método proposto foi comparado com o obtido por HPLC-UV e com espectrofotometria de absorção molecular (ambos após derivação química do analito com o-ftalaldeído ou com ninhidrina). O método desenvolvido neste trabalho foi aplicado em formulações farmacêuticas injetáveis e em amostras simuladas, com recuperações próximas de 100 por cento. Para a determinação do lapachol foi utilizada a SWV usando um eletrodo de grafite-epoxi feito no laboratório. O meio eletrolítico utilizado foi uma solução aquosa contendo o surfactante catiônico CTAB (1,2 × 10-4 mol L-1), tampão fosfato (4,0 × 10-2 mol L-1, pH 6,0) e KNO3 (1,0 mol L-1). O surfactante catiônico melhorou a difusão e a interação do analito com o eletrodo, produzindo um processo reversível que melhorou a corrente total medida pelo sistema. O sinal do lapachol foi medido a -470 mV, após pré-concentração a 400 mV durante 140 s, usando frequência de 30 Hz, amplitude de pulso de 40 mV com passo de potencial de 20 mV. O limite instrumental de detecção foi de 0,029 mg L-1 e a faixa linear dinâmica abrangeu duas ordens de grandeza. Na presença de beta-lapachona (isômero estrutural), a seletividade foi obtida pela derivada de primeira ordem do sinal SWV. A determinação do lapachol em extrato etanólico de Tabebuia impetiginosa foi realizada após uma s / [en] The present work aimed to develop electroanalytical methods using graphite-epoxy and boron-doped diamond electrodes for the determination of the antibiotic gentamicin sulfate, substances of biological interest (naphthoquinones) and the pesticide trifloxystrobin. Boron doped diamond (BDD) features attractive electrochemical detection characteristics that are useful in analytical applications based on voltammetry and amperometry. It has a wide potential window in aqueous solutions that allowed the quantification of the fungicide trifloxystrobin (in +1744 mV versus Ag/AgCl(KClsat)) by square-wave voltammetry (SWV) in Britton-Robinson buffer (0.04 mol L-1, pH 4.00) / acetonitrile 70/30 percent v/v electrolyte medium. The analyte addition curve was constructed using 40 mV pulse amplitude, 30 Hz frequency, 20 mV step potential. The instrumental limit of detection was 1.4 × 10-7 mol L-1 and the linear dynamic range covered three orders of magnitude (from 10-7 to 10-5 mol L-1). Recoveries of about 80 percent in orange juice samples and 92.4 to 104.0 percent in water samples. A study to evaluate potential interferences was carried out in the presence of other fungicides. Diagnostic studies indicated that the oxidation of trifloxystrobin in aqueous medium, on the surface of the BDD, is irreversible, involving two steps, each of them with two electrons. UV degradation of trifloxystrobin was studied using the proposed voltammetric method and degradation kinetics with a half-life of 1.07 min was established. Batch injection amperometry (BIA) with BDD (operating at +2000 mV) was used to develop a method for the determination of gentamicin sulfate. Solid phase extraction (SPE) using a cartridge packed with a molecular imprinting polymer made with kanamycin as the template molecule, was used to improve the selectivity. The acetate buffer (0.01 mol L-1; pH 4.40) was used as supporting electrolyte. Oxidation of gentamicin sulfate occurred right after the analyte (60 microlitres solution) comes into contact with the surface of the electrode. The SPE cartridge packed with MIP was very efficient in retaining the analyte when compared to the one packed with the corresponding non-printed polymer (NIP). The analytical curve was linear, the instrument accuracy was less than 3 percent and the limit of quantification was 2.7 × 10-6 mol L-1. The analytical frequency was 90 measurements h-1 and no memory effect was observed between sequential measurements. The performance of the proposed method was compared with that obtained by HPLC-UV and molecular absorption spectrophotometry (both after chemical derivatization of the analyte with o-phthalaldehyde or with ninhydrin). The method was applied in injectable pharmaceutical formulations and in simulated samples, with recoveries close to 100 percent. For the determination of lapachol SWV was employed using a graphite-epoxy electrode made in the laboratory. The electrolytic medium used was an aqueous solution containing the cationic surfactant CTAB (1.2 × 10-4 mol L-1), phosphate buffer (4.0 × 10 -2 mol L-1, pH 6.00) and KNO3 (1.0 mol L-1). The cationic surfactant improved the diffusion and interaction of the analyte with the electrode, producing a reversible process that improved the total current measured by the system. The lapachol signal was measured at -470 mV after preconcentration at 400 mV for 140 s, using a 30 Hz frequency, 40 mV pulse amplitude with a step potential of 20 mV. The instrumental limit of detection was 0.029 mg L-1 and the dynamic linear range comprised two orders of magnitude. In the presence of beta-lapachone (structural isomer), the selectivity was obtained by the first-order derivative of the SWV signal. The determination of lapachol in ethanolic extract of Tabebuia impetiginosa was carried out after a simple separation of the analyte by thin layer chromatography. The results are statistically similar (with 95 percent confidence level) with those performed through HPLC-UV. Studies for the simultaneous determination of beta-lapachone and alpha-lapachone were also successful
|
9 |
[pt] DESENVOLVIMENTO DE MÉTODOS ELETROANALÍTICOS PARA A DETERMINAÇÃO DE FUNGICIDAS DA CLASSE DAS ESTROBILURINAS UTILIZANDO ELETRODOS DE FILME DE BISMUTO E DE DIAMANTE DOPADO COM BORO / [en] ELECTROANALYTICAL METHODS FOR THE DETERMINATION OF FUNGICIDES OF THE STROBILURIN CLASS USING THE BISMUTH FILM ELECTRODE AND THE BORON-DOPED DIAMOND29 November 2021 (has links)
[pt] O monitoramento de resíduos de pesticidas é extremamente importante e isso se dá pelo emprego de métodos analíticos adequados para avaliar a sua presença em compartimentos ambientais e a contaminação de alimentos com esses agentes. Os pesticidas da classe das estrobilurinas vêm se tornando os mais utilizados para o controle dos fungos e apesar da propagada baixa toxicidade e rápida degradação, estudos devem ser realizados para averiguar o nível de exposição dos humanos a tais substâncias. Estudos com as estrobilurinas (azoxistrobina, cresoxim-metílico, dimoxistrobina, fluoxastrobina, picoxistrobina, piraclostrobina e trifloxistrobina) utilizando o eletrodo de trabalho de diamante dopado com boro (DDB) foram realizados por voltametria cíclica e voltametria de onda quadrada (SWV) e indicaram dois picos de oxidação (varredura anódica) em potenciais distintos. Dados experimentais e da literatura permitiram supor que para cresoxim-metílico e dimoxistrobina o primeiro pico de oxidação tem um mecanismo irreversível com 2 mols de elétrons por mol de analito e o segundo pico de oxidação com mecanismo quase-reversível. Para picoxistrobina, os dois picos de oxidação são irreversíveis com 2 mols de elétrons por mol de analito. Para piraclostrobina os dois picos de oxidação são quase-reversíveis. Métodos por SWV foram desenvolvidos utilizando DDB para determinação de cresoxim-metílico, picoxistrobina e piraclostrobina. As varreduras anódicas foram feitas entre: +1000 a +1750 mV (medição do sinal em +1420 mV) para cresoxim-metílico; +1100 a +1800 mV (medição do sinal em +1450 mV) para picoxistrobina e +1050 a +2100 mV (medição do sinal em +1280 mV) para piraclostrobina. O eletrólito suporte foi tampão acetato (0,05 mol L-1; pH 4,0) para a determinação de cresoxim-metílico e piraclostrobina e tampão Britton-Robinson (0,04 mol L-1; pH 2,0) para a picoxistrobina. Resposta analítica linear para cresoxim-metílico foi do limite de quantificação, LOQ, (0,9 umol L-1 ou 0,27 mg L-1) até 34,0 μmol L-1 com R2> 0,999. Para picoxistrobina a faixa linear foi do LOQ (0,7 mol L-1 ou 0,25 mg L-1) a 20,0 μmol L-1 (R2> 0,994). Já para piraclostrobina a faixa linear foi do LOQ (0,8 μmol L-1 ou 0,32 mg L-1) a 20,0 μmol L-1 (R2> 0,991). Os métodos foram aplicados na análise de amostras de suco de uva (cresoxim-metílico e piraclostrobina) e águas naturais (picoxistrobina e piraclostrobina). As recuperações variaram de 91,6 a 105,3 porcento dependendo da estrobilurina. Os métodos propostos foram comparados com cromatografia líquida de alta eficiência (HPLC) e os resultados não indicaram diferença significativa entre eles. Amostras de suco de uva comercial foram fortificadas na concentração do limite máximo permitido pela legislação, sendo o analito pré-concentrado dez vezes por extração em fase sólida (SPE), com cartucho C18, obtendo-se recuperações entre 95,4 a 99,5 porcento dependendo da estrobilurina. Dados experimentais indicaram que não houve degradação térmica significativa para as estrobilurinas em soluções aquecidas a 60 °C durante 2 h. Não foi possível propor um perfil simples de cinética para degradação por exposição ao UV para o cresoxim-metílico. Para picoxistrobina e piraclostrobina, um modelo de cinética de segunda ordem pode ser atribuído para a degradação por exposição ao UV. Métodos com a abordagem de análise por injeção em batelada (BIA) com detecção amperométrica no eletrodo de DDB foram desenvolvidos para a determinação de dimoxistrobina e picoxistrobina em águas naturais. Os métodos apresentaram uma frequência de análise de 180 injeções h-1 (dimoxistrobina) e 108 injeções h-1 (picoxistrobina), e precisão satisfatória (abaixo de 5 porcento). Resposta analítica linear para dimoxistrobina foi de 1,3 μmol L-1 (0,41 mg L-1) a 60,0 μmol L-1 com R2> 0,999 e para picoxistrobina foi de 5,3 μmol L-1 (1,95 mg L-1) e 100,0 μmol L-1 (R2> 0,999). Valores de recuperação obtidos nas amostras de águas naturais fortificadas com os analitos foram entre 80,2 e 105,6 porcento dependendo do analito. Além disso, uma estratégia simples para detectar a presença de moléculas interferentes em amostras de água com base na amperometria de múltiplos pulsos foi apresentada com sucesso. Um método voltamétrico de redissolução anódica por pulso diferencial usando o eletrodo de filme de bismuto foi desenvolvido para a determinação de picoxistrobina em amostras de urina e águas naturais. O ciclo de medição começou com a aplicação de um potencial de deposição de -700 mV durante 60 s para a formação in situ do filme Bi e a acumulação do analito. A varredura de potencial foi no sentido anódico entre +790 e +1050 mV. Provou-se que, nas condições empregadas, uma quantidade significativa de filme de Bi ainda está presente na superfície do eletrodo de carbono vítreo após a varredura anódica, com uma velocidade de 40 mV s-1. A presença de Bi é fundamental para a acumulação e oxidação de picoxistrobina (pico máximo em +954 mV) durante redissolução. O eletrólito suporte foi HCl 1,0 mol L-1. Um procedimento de limpeza foi desenvolvido a fim de minimizar o efeito de memória. Resposta analítica linear para picoxistrobina foi observada de 2,8 μmol L-1 (1,0 mg L-1) até 19,0 μmol L-1 (R2> 0,994). SPE permitiu a pré-concentração do analito e eliminou as interferências nas amostras de urina. Recuperações de 89,3 a 104,8 porcento foram encontradas e a interferência das outras estrobilurinas ao método foram avaliadas. O método proposto foi comparado com HPLC e os resultados indicam que não houve diferença significativa entre eles. / [en] The monitoring of pesticide residues is extremely important and this is accomplished by the use of methods to assess their presence in environmental compartments and in food stuff contaminated with these agents. The pesticides of the strobilurin class have become the most widely used for fungi control and despite the low toxicity and rapid degradation, studies should be conducted to ascertain the level of human exposure to such substances. Studies with the strobilurin pesticides (azoxystrobin, kresoxim-methyl, dimoxystrobin, fluoxastrobin, picoxystrobin, pyraclostrobin and trifloxystrobin) were performed by cyclic voltammetry and square wave voltammetry (SWV) using the boron-doped diamond (BDD). Results indicated two oxidation peaks (anodic scan) at different potentials. Based on experimental data and literature it is assumed that for kresoxim-methyl and dimoxystrobin the first oxidation peak comes from an irreversible mechanism involvieng 2 moles of electrons per mole of analyte while the second oxidation peak is from a quasi-reversible mechanism. For picoxystrobin, the two oxidation peaks are from irreversible processes involving 2 moles of electrons per mole of analyte. For pyraclostrobin, the two oxidation peaks are from quasi-reversible processes. SWV methods were developed for determination of kresoxim-methyl, pyraclostrobin and picoxystrobin using the BDD. The anodic scans were made between: +1000 to +1750 mV (measured signal at +1420 mV) for kresoxim-methyl; +1100 to +1800 mV (measured signal at +1450 mV) for picoxystrobin and +1050 to +2100 mV (measured signal at +1280 mV) for pyraclostrobin. The supporting electrolyte was acetate buffer (0.05 mol L-1, pH 4.0) for the determination of kresoxim-methyl and pyraclostrobin and Britton-Robinson buffer (0.04 mol L-1, pH 2.0) to picoxystrobin. Linear analytical response to kresoxim-methyl was the limit of quantification, LOQ (0.9 μmol L-1 or 0.27 mg L-1) to 34.0 μmol L-1 with R2> 0.999. For picoxystrobin the linear range was from LOQ (0.7 μmol L-1 or 0.25 mg L-1) to 20.0 μmol L-1 (R2> 0.994). Pyraclostrobin presented linear range from LOQ (0.8 μmol L-1 and 0.32 mg L-1) to 20.0 μmol L-1 (R2> 0.991). The methods were applied in the analysis of grape juice samples (kresoxim-methyl and pyraclostrobin) and in natural waters (picoxystrobin and pyraclostrobin). The recoveries ranged from 91.6 to 105.3 percent depending on the strobilurin. The results obtained with the proposed methods were compared with the ones achieved by high performance liquid chromatography (HPLC) indicating no significant difference between them. Commercial grape juice samples were fortified at the maximum concentration allowed legislation and the analytes were pre-concentrated (tenfold) by solid phase extraction (SPE) in a C18 cartridge analyte Recoveries were between 95.4 to 99.5 percent depending on the strobilurin. Experimental data indicate that no significant thermal degradation occurred for strobilurins in solutions heated at 60 °C for 2 h. It was not possible to propose a simple kinetic profile for the UV degradation of kresoxim-methyl. However, for picoxystrobin and pyraclostrobin, a second-order kinetic model explains their degradation profile. Methods using batch injection analysis (BIA) with amperometric detection on a BDD electrode have also been developed for the determination of picoxystrobin and dimoxystrobin in natural waters. The methods showed an analytical frequency of 180 injections h-1 (dimoxystrobin) and 108 injections h-1 (picoxystrobin) and satisfactory precision (less than 5 percent). Linear analytical response for dimoxystrobin was from 1.3 μmol L-1 (0.41 mg L-1) to 60.0 μmol L-1 with R2> 0.999 while for picoxystrobin, the range was from 5.3 μmol L-1 (1.95 mg L-1) to 100.0 μmol L-1 (R2> 0.999). Recoveries achieved for analyte fortified water samples were between 80.2 and 105.6 percent depending on the analyte. In addition, a simple strategy to detect the presence of interfering molecules in water samples, based on multiple pulse amperometry, was successfully introduced. A stripping anodic differential pulse voltammetric method using the bismuth film electrode was developed for the determination picoxystrobin in urine samples and in natural waters. The measurement cycle started with the application of a deposition potential of -700 mV for 60 s for the in situ formation of the Bi film and for the accumulation of analyte. The potential was scanned in the anodic direction between +790 and +1050 mV. It has been proven that, under the conditions employed, a significant amount of Bi film is still present on the surface of the glassy carbon electrode after anodic sweep, with a speed of 40 mV s-1. The presence of Bi is essential for the accumulation and oxidation of picoxystrobin (maximum peak at +954 mV) during stripping. The supporting electrolyte was HCl 1.0 mol L-1. A cleaning procedure was developed to minimize the memory effect. Analytical linear response was observed for picoxystrobin from 2.8 μmol L-1 (1.0 mg L-1) to 19.0 μmol L-1 (R2> 0.994). SPE allowed the pre-concentration of the analyte and elimination of interferences from urine samples. Recoveries from 89.3 to 104.8 percent were found. Interference of other strobilurin the method were evaluated. The proposed method was compared with HPLC and the results indicate no significant difference between them.
|
Page generated in 0.0338 seconds